Does robotic-assisted computer navigation improve acetabular cup positioning in total hip arthroplasty for Crowe
Wei Chai1,2,3, Chi Xu1,2,3, Ren-Wen Guo1
1Senior Department of Orthopedics, the Fourth Medical Center of Chinese PLA General Hospital, Beijing, China.
Aims:
Total hip arthroplasty (THA) in patients with hip-dislocation dysplasia remains challenging. This study aims to evaluate whether these patients may benefit from robotic-assisted techniques.
Methods:
We reviewed 135 THAs (108 conventional THAs and 27 robotic-assisted THAs) for Crowe type III or IV from January 2017 to August 2019 in our institution. Robotic-assisted THAs were matched with conventional THAs at a 1:1 ratio (27 hips each group) using propensity score matching. The accuracy of cup positioning and clinical outcomes were compared between groups.
Results:
The inclination of the cup for conventional THAs and robotic THAs was 42.1 ± 5.7 and 41.3 ± 4.6 (p = 0.574), respectively. The anteversion of the cup for conventional THAs was significantly greater than that of robotic THAs (29.5 ± 8.1 and 18.0 ± 4.6; p < 0.001), respectively. The ratio of the acetabular cup in the Lewinnek safe zone was 37% (10/27) in conventional THAs and 96.3% (26/27) in robotic THAs (p < 0.001). Robotic THAs did not achieve better leg length discrepancy than that of conventional THAs (- 0.4 ± 10.9 mm vs. 0.4 ± 8.8 mm, p = 0.774). There was no difference in Harris Hip Score and WOMAC Osteoarthritis index between groups at the 2-year follow-up. No dislocation occurred in all cases at the final follow-up.
Conclusion:
Robotic-assisted THA for patients with high dislocation improves the accuracy of the implantation of the acetabular component with respect to safe zone.