Understanding and correcting wavenumber error in interference pattern structured illumination imaging

Optics Express
|February 24, 2022
PubMed

Related Concept Videos

Phase Contrast and Differential Interference Contrast Microscopy01:26

Phase Contrast and Differential Interference Contrast Microscopy

Phase-Contrast Microscopes
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
11.1K
Interference: Path Lengths01:10

Interference: Path Lengths

Consider two sources of sound, that may or may not be in phase, emitting waves at a single frequency, and consider the frequencies to be the same.
Two special sources may be considered when they are in phase. This can be easily achieved by feeding the two sources from the same source. An example would be synchronizing the two speakers by feeding them with the same source, such as the sound waves produced by a tuning fork. This setup ensures that the two sources have the same frequency and are...
1.4K
NMR Spectrometers: Resolution and Error Correction01:14

NMR Spectrometers: Resolution and Error Correction

When magnetic nuclei in a sample achieve resonance and undergo relaxation, the signal detected in NMR is an approximately exponential free induction decay. Fourier transform of an exponential decay yields a Lorentzian peak in the frequency domain. Lorentzian peaks in an NMR spectrum are defined by their amplitude, full width at half maximum, and position, where the peak width is governed by the spin-spin relaxation time alone. In real experiments, however, the applied magnetic field is rendered...
791