A real-world experience of atrioventricular synchronous pacing with leadless ventricular pacemakers
Gurukripa N Kowlgi1, Andrew S Tseng1, Nathan D Tempel1
1Department of Cardiovascular Medicine, Mayo Clinic, Rochester, Minnesota, USA.
Aims:
The MicraTM transcatheter pacing system (TPS) (Medtronic) is the only leadless pacemaker that promotes atrioventricular (AV) synchrony via accelerometer-based atrial sensing. Data regarding the real-world experience with this novel system are scarce. We sought to characterize patients undergoing MicraTM -AV implants, describe percentage AV synchrony achieved, and analyze the causes for suboptimal AV synchrony.
Methods:
In this retrospective cohort study, electronic medical records from 56 consecutive patients undergoing MicraTM -AV implants at the Mayo Clinic sites in Minnesota, Florida, and Arizona with a minimum follow-up of 3 months were reviewed. Demographic data, comorbidities, echocardiographic data, and clinical outcomes were compared among patients with and without atrial synchronous ventricular pacing (AsVP) ≥ 70%.
Results:
Sixty-five percent of patients achieved AsVP ≥ 70%. Patients with adequate AsVP had smaller body mass indices, a lower proportion of congestive heart failure, and prior cardiac surgery. Echocardiographic parameters and procedural characteristics were similar across the two groups. Active device troubleshooting was associated with higher AsVP. The likely reasons for low AsVP were small A4-wave amplitude, high ventricular pacing burden, and inadequate device reprogramming. Importantly, in patients with low AsVP, subjective clinical worsening was not noted during follow-up.
Conclusion:
With the increasing popularity of leadless pacemakers, it is paramount for device implanting teams to be familiar with common predictors of AV synchrony and troubleshooting with MicraTM -AV devices.
Related Concept Videos
Electrophysiology of Normal Cardiac Rhythm
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization II: Right Heart Catheterization
Cardiopulmonary Resuscitation III: AED Use
Cardiac Catheterization I: Pre-Procedure Overview