Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Gas Chromatography–Mass Spectrometry (GC–MS)01:14

Gas Chromatography–Mass Spectrometry (GC–MS)

4.9K
Gas chromatography–mass spectrometry (GC–MS) is the combination of analytical techniques of gas chromatography and mass spectrometry in a single instrument for analyzing a mixture of compounds. The gas chromatograph separates the compounds in the mixture, and the mass spectrometer analyzes each compound separately to determine the molecular masses and molecular structures.
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
4.9K
Gas Chromatography: Introduction01:13

Gas Chromatography: Introduction

2.4K
Gas chromatography (GC) is a technique for separating and analyzing volatile compounds in a sample. Its primary purpose is to identify and quantify components in complex mixtures, making it essential in fields such as environmental analysis, pharmaceuticals, and petrochemicals. GC is also called vapor-phase chromatography (VPC) or gas-liquid partition chromatography (GLPC).
In GC,  a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
2.4K
Gas Chromatography: Types of Columns and Stationary Phases01:17

Gas Chromatography: Types of Columns and Stationary Phases

1.1K
Gas chromatography (GC) relies on stationary phases to separate and analyze components in a sample. There are two main types of stationary phases: liquid and solid. Liquid stationary phases are non-volatile, thermally stable, and chemically inert liquids coated onto the column. Solid stationary phases are particles of adsorbent material, such as silica gel or molecular sieves.
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
1.1K
Gas Chromatography: Sample Injection Systems01:08

Gas Chromatography: Sample Injection Systems

623
In gas chromatography, the sample is introduced as a vapor plug into the carrier gas stream for high efficiency and resolution. A microsyringe injects the sample solution into a heated sample port, vaporizing it and mixing it with the carrier gas. This process is important to ensure the sample is properly prepared for analysis. Thermally sensitive samples can be injected directly into the column and volatilized by slowly increasing the column temperature.
Two primary injection methods are used...
623
Gas Chromatography: Overview of Detectors01:13

Gas Chromatography: Overview of Detectors

875
Detectors in gas chromatography (GC) help identify and quantify the components of a mixture by translating chemical properties into measurable signals, which are displayed on a chromatogram. Detectors can be categorized into two main types: destructive and non-destructive.
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
875
Gas Chromatography: Types of Detectors-II01:19

Gas Chromatography: Types of Detectors-II

543
In gas chromatography, different detectors are employed to meet specific analytical needs. These detectors are often categorized based on their detection mechanisms and the types of compounds they are best suited to analyze. Thermal Conductivity Detectors (TCD), Flame Ionization Detectors (FID), and Electron Capture Detectors (ECD) represent common categories, each with unique operating principles and applications. However, beyond these, several other detectors are designed for more specialized...
543

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

A fast temperature-programmed second-dimension column for comprehensive two-dimensional gas chromatography.

Analytical and bioanalytical chemistry·2022
Same author

Radiation Safety Considerations in the Treatment of Canine Skeletal Conditions Using 153Sm, 90Y, and 117mSn.

Health physics·2020
Same author

A comprehensive two-dimensional gas chromatography valve modulation method using hold-release primary column flow for long secondary separation time with 100% transfer.

Journal of chromatography. A·2018
Same author

Compact Ultrahigh-Pressure Nanoflow Capillary Liquid Chromatograph.

Analytical chemistry·2017
Same author

Hand-portable liquid chromatographic instrumentation.

Journal of chromatography. A·2015
Same author

Hand-Portable Gradient Capillary Liquid Chromatography Pumping System.

Analytical chemistry·2015

Related Experiment Video

Updated: Sep 30, 2025

Chromatographic Fingerprinting by Template Matching for Data Collected by Comprehensive Two-Dimensional Gas Chromatography
10:14

Chromatographic Fingerprinting by Template Matching for Data Collected by Comprehensive Two-Dimensional Gas Chromatography

Published on: September 2, 2020

5.1K

Comprehensive two-dimensional gas chromatography using direct flow modulation to extend the secondary dimension

Huamin Cai1, Stanley D Stearns1

  • 1Valco Instruments Co. Inc.

Journal of Chromatography. A
|March 16, 2022
PubMed
Summary

A novel direct flow modulator enhances gas chromatography (GC) separations by enabling longer secondary retention times. This method improves peak resolution and is applicable to complex samples like gasoline.

Keywords:
Comprehensive two-dimensional gas chromatographyGCFlow modulatorGC×Gasoline analysisLong secondary separation timeReference gas oil standard analysis

More Related Videos

Qualitative Characterization of the Aqueous Fraction from Hydrothermal Liquefaction of Algae Using 2D Gas Chromatography with Time-of-flight Mass Spectrometry
11:44

Qualitative Characterization of the Aqueous Fraction from Hydrothermal Liquefaction of Algae Using 2D Gas Chromatography with Time-of-flight Mass Spectrometry

Published on: March 6, 2016

9.5K
Measurement of H2S in Crude Oil and Crude Oil Headspace Using Multidimensional Gas Chromatography, Deans Switching and Sulfur-selective Detection
08:37

Measurement of H2S in Crude Oil and Crude Oil Headspace Using Multidimensional Gas Chromatography, Deans Switching and Sulfur-selective Detection

Published on: December 10, 2015

19.4K

Related Experiment Videos

Last Updated: Sep 30, 2025

Chromatographic Fingerprinting by Template Matching for Data Collected by Comprehensive Two-Dimensional Gas Chromatography
10:14

Chromatographic Fingerprinting by Template Matching for Data Collected by Comprehensive Two-Dimensional Gas Chromatography

Published on: September 2, 2020

5.1K
Qualitative Characterization of the Aqueous Fraction from Hydrothermal Liquefaction of Algae Using 2D Gas Chromatography with Time-of-flight Mass Spectrometry
11:44

Qualitative Characterization of the Aqueous Fraction from Hydrothermal Liquefaction of Algae Using 2D Gas Chromatography with Time-of-flight Mass Spectrometry

Published on: March 6, 2016

9.5K
Measurement of H2S in Crude Oil and Crude Oil Headspace Using Multidimensional Gas Chromatography, Deans Switching and Sulfur-selective Detection
08:37

Measurement of H2S in Crude Oil and Crude Oil Headspace Using Multidimensional Gas Chromatography, Deans Switching and Sulfur-selective Detection

Published on: December 10, 2015

19.4K

Area of Science:

  • Analytical Chemistry
  • Chromatography

Background:

  • Two-dimensional gas chromatography (GC×GC) offers enhanced separation power for complex mixtures.
  • Optimizing the interface between the primary and secondary columns is crucial for effective GC×GC analysis.

Purpose of the Study:

  • To develop and evaluate a simple direct flow modulator for GC×GC.
  • To investigate factors influencing secondary dimension separation and peak width.
  • To demonstrate the modulator's applicability to real-world samples.

Main Methods:

  • A direct flow modulator was constructed using a tee connector to introduce pulsed flow between GC columns.
  • The effects of modulation pulse width, period, and pressure differential on secondary peak width (²Wb) were studied.
  • Different tee connector sizes (360 µm, 1/32″, 1/16″) were tested.

Main Results:

  • Achieved a 120 s secondary separation time, sufficient for longer secondary columns and temperature ramping.
  • Modulation parameters significantly impact ²Wb by affecting primary effluent loading.
  • The 360 µm tee connector provided the best performance with reduced peak tailing and narrower ²Wb.

Conclusions:

  • The developed direct flow modulator effectively improves GC×GC separation.
  • Optimization of modulation parameters and tee connector size is key to achieving narrow peaks and high resolution.
  • The method is suitable for analyzing complex samples such as gasoline and gas oil.