Comparison Study of the Two Biometers Based on Swept-Source Optical Coherence Tomography Technology.
Jing Dong1, Jinhan Yao2, Shuimiao Chang2
1Department of Ophthalmology, First Hospital of Shanxi Medical University, Taiyuan 030001, China.
Diagnostics (Basel, Switzerland)
|March 25, 2022
Summary
Two swept-source OCT biometers showed significant differences in most measurements, including IOL power, except for axial length. This highlights potential variability in cataract surgery planning.
Area of Science:
- Ophthalmology
- Medical Devices
- Optical Coherence Tomography
Background:
- Accurate biometry is crucial for successful intraocular lens (IOL) power calculation in cataract surgery.
- Swept-source optical coherence tomography (SS-OCT) biometers are widely used for ophthalmic measurements.
Purpose of the Study:
- To compare key biometric parameters measured by two different SS-OCT biometers: ANTERION (biometer A) and IOLMaster 700 (biometer B).
- To assess potential differences in intraocular lens (IOL) power calculations derived from these two devices.
Main Methods:
- A prospective observational comparative case series study involving 179 eyes undergoing cataract surgery.
- Measurements included axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), keratometry (Kf, Ks, Km), astigmatism, white-to-white (WTW) distance, and IOL power.
- Paired t-tests and McNemar tests were used for statistical comparison.
Main Results:
- No significant difference was found in axial length (AL) or astigmatism magnitude between the two biometers.
- Significant differences were observed in CCT, ACD, LT, keratometry (Kf, Ks, Km), WTW distance, astigmatism axis, and IOL power.
- Approximately 65.4% of eyes showed a ≥0.5-D difference in IOL power between devices.
Conclusions:
- The ANTERION and IOLMaster 700 SS-OCT biometers exhibit significant measurement differences for most ocular parameters, excluding AL.
- These discrepancies, particularly in IOL power, may impact refractive outcomes following cataract surgery and warrant careful consideration.