AutoInFocus, a new paradigm for ultrasound-guided spine intervention: a multi-platform validation study
Keshuai Xu1, Baichuan Jiang1, Abhay Moghekar2
1Department of Computer Science, Johns Hopkins University, Baltimore, 21218, MD, USA.
Purpose:
Ultrasound-guided spine interventions often suffer from the insufficient visualization of key anatomical structures due to the complex shapes of the self-shadowing vertebrae. Therefore, we propose an ultrasound imaging paradigm, AutoInFocus (automatic insonification optimization with controlled ultrasound), to improve the key structure visibility.
Methods:
A phased-array probe is used in conjunction with a motion platform to image a controlled workspace, and the resulting images from multiple insonification angles are combined to reveal the target anatomy. This idea is first evaluated in simulation and then realized as a robotic platform and a miniaturized patch device. A spine phantom (CIRS) and its CT scan were used in the evaluation experiments to quantitatively and qualitatively analyze the advantages of the proposed method over the traditional approach.
Results:
We showed in simulation that the proposed system setup increased the visibility of interspinous space boundary, a key feature for lumbar puncture guidance, from 44.13 to 67.73% on average, and the 3D spine surface coverage from 14.31 to 35.87%, compared to traditional imaging setup. We also demonstrated the feasibility of both robotic and patch-based realizations in a spine phantom study.
Conclusion:
This work lays the foundation for a new imaging paradigm that leverages redundant and controlled insonification to allow for imaging optimization of the complex vertebrae anatomy, making it possible for high-quality visualization of key anatomies during ultrasound-guided spine interventions.