Related Experiment Video
Updated: Sep 26, 2025

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Staff eye lens dose in interventional radiology and cardiology in Finland
Antti Pekkarinen1, Carita Lindholm2, Mika Kortesniemi3
1Radiation and Nuclear Safety Authority (STUK), POB 14, Helsinki 00811, Finland; HUS Diagnostic Center, Helsinki University and Helsinki University Hospital, Haartmaninkatu 4, 00290 Helsinki, Finland; Department of Physics, University of Helsinki, Helsinki, Finland; Department of Medical Physics, Kymsote, Kymenlaakso Central Hospital, Kotkantie 41, 48210 Kotka, Finland.
Purpose:
The aim of this study was to investigate the eye lens and whole-body radiation doses to interventional radiology and cardiology staff in two Finnish hospitals.
Methods:
Simultaneous measurements of personal dose equivalent quantities Hp(3) and Hp(10) were conducted in clinical conditions during different radiological and cardiological interventional procedures. In order to study the feasibility to estimate eye lens dose with Hp(10) measured over the protective apron or thyroid shield, the ratio between measured Hp(3) and Hp(10) was investigated.
Results And Conclusions:
Applying the obtained ratio on Hp(10) records from national dose register showed that only a small number of interventional radiologists and cardiologists in Finland may exceed eye lens equivalent dose levels of 20 mSv per year or 100 mSv in five consecutive years, but likely do not exceed 50 mSv in a single year. For the most Finnish interventionalists, the eye lens dose is well below 10 mSv per year. Nurses and radiographers assisting in interventions are, on average, less exposed than interventionalists, and will not exceed 20 mSv per year. Based on our results, Hp(10) measured over the protective apron or thyroid shield provides a conservative estimate of the eye lens dose for interventional radiologists and cardiologists, provided that appropriate protective glasses are used.
More Related Videos
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
06:51Dosimetry for Cell Irradiation using Orthovoltage 40-300 kV X-Ray Facilities
Published on: February 20, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Renal Failure: Dose Adjustments
Reduced renal clearance and elimination rate are common outcomes of renal impairment. These alterations lead to a prolonged elimination half-life and an altered apparent volume of distribution for drugs. As a result, dosage adjustments are typically necessary to maintain optimal drug levels in the body.
However, dosage adjustments...