Related Experiment Video
Updated: Sep 24, 2025

Author Spotlight: Development of a Smartphone-Enhanced Paper-Based Device for Rapid Dengue NS1 Detection
Published on: January 26, 2024
New, inexpensive and simple 3D printable device for nephelometric and fluorimetric determination based on smartphone
Ezequiel Vidal1, Anabela S Lorenzetti1, Miguel Ángel Aguirre2
1INQUISUR, Departamento de Química, Universidad Nacional del Sur (UNS)-CONICET Av. Alem 1253 8000 Bahía Blanca Argentina ezquielvidal.estudio@gmail.com claudia.domini@uns.edu.ar +54 291 4595101 int 3558.
Abstract:
A new, inexpensive and easy to use 3D printable device was developed for nephelometric and fluorimetric determination. Its applicability was tested for the quantification of quinine in tonic drinks and sulfate in natural water with good analytical accuracy. In this way, sulfate determination was carried out by nephelometry using a red LED, while quinine was determined using a blue LED by fluorimetry. A smartphone camera was used to take the pictures and afterwards transform them into the RGB color space using the software ImageJ by a personal computer. The linear range was 2.0-50.0 mg L-1 for sulfate with a LOD of 0.13 mg L-1, and the corresponding quantification limit (LOQ) was 0.43 mg L-1. The linear range for quinine was from 0.42 to 3.10 mg L-1. The LOD and LOQ were 0.11 mg L-1 and 0.38 mg L-1, respectively.