Miniaturized 3D printed electrochemical platform with optimized Fibrous carbon electrode for non-interfering
Jaligam Murali Mohan1, Khairunnisa Amreen2, Arshad Javed1
1Department of Mechanical Engineering, Birla Institute of Technology and Science (BITS) Pilani, Hyderabad Campus, Hyderabad, 500078, India.
Abstract:
3D printing technology based electrochemical device can provide ease of fabrication, cost effectiveness, rapid detection and lower limit of detection. Herein, a novel, customized, portable and inexpensive 3D printed electrochemical device, has been presented. Fibrous carbon Toray paper, deposited with gold nanoparticles through electrodeposition, used as a working electrode which Further device was tested with 1 mM sodium hypochlorite using cyclic voltammetry (CV) and square wave voltammetry (SWV) in 0.1 M PBS. Hypochlorite has a pivotal role in supporting the growing chemical and paper industries and finds diverse uses in several clinical applications. It is primarily used for disinfecting food, water and surfaces. The scan rate study was carried out from 20 mVs-1 to 250 mVs-1 using cyclic voltammetry technique. The diffusion coefficient obtained from scan rate effect was 1.39 × 10-6 cm2s-1. The concentration range was evaluated with SWV technique, in a linear range of 0.6 μM-40 μM, with a detection limit of 0.7 μM. The device was further analyzed to ensure non-interference from co-existing chemicals like sodium chloride, potassium nitrate, sodium carbonate, sodium nitrite. Real sample analysis was done with sea, artificial sea and tap water with impressive recovery values. In summary, the developed working electrode can be customized and modified based on testing analyte; thus, the proposed device can be used for various other biochemical analytes.