Software for Analysis of Heart Rate and Blood Pressure Time-series Data from the Valsalva Maneuver
Published on: June 27, 2025
Automated extraction of anorectal pressures from high-resolution manometry reports
Mohamed ElWazir1, Misha Gautam2, Rahul Mishra2
1Department of Cardiovascular Medicine, Suez Canal University, Ismailia, Egypt.
Background:
During high-resolution anorectal manometry (HR_ARM), the raw data are converted into software-derived summary variables (e.g., rectoanal gradient during evacuation) that capture only a snapshot of the data collected during HR_ARM and are less useful than newer indices, which are also derived from the raw data, for diagnosing defecatory disorders (DD). However, it is cumbersome and time-consuming to extract raw data from the program and calculate such indices. This study aimed to devise a user-friendly approach to extract anorectal pressures measured with HR_ARM.
Methods:
Anorectal pressures at rest, during anal squeeze, and evacuation were measured with HR_ARM in 180 women, among whom 60 each were healthy, had DD, or fecal incontinence. A reverse engineering approach extracted pressure values from color images in HR_ARM reports. The summary variables generated by the software and a reverse engineering approach were compared with Lin's concordance correlation coefficient (CCC), paired t-tests, and Bland-Altman's tests.
Results:
Anorectal pressures summarized by the software and a reverse engineering method were highly concordant for anal resting (CCC [95% CI], 0.98 [0.97, 0.99]) and squeeze pressures (0.99 [0.99, 0.99) and the rectoanal gradient during evacuation with an empty (0.98 [0.97, 0.98]) and a filled balloon (0.99 [0.99, 0.99]). For most variables, the paired t and Bland-Altman comparisons were not significant.
Conclusions:
Anorectal pressures can be accurately determined from pressure topography images in HR_ARM reports. In future, this reverse engineering approach can be harnessed to compile large HR_ARM datasets across centers and to uncover newer, potentially more useful summary rectoanal pressure variables.
More Related Videos
Related Concept Videos
Assessment of the Rectum and Anus
Rectal Inspection
Begin by inspecting the perianal and anal areas for color, texture, rashes,...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...