Organs' absorbed dose and comparison of different methods for effective dose calculation in computed tomography of
Christos P Dimitroukas1,2, Vasileios I Metaxas1, Fotios O Efthymiou1
1Department of Medical Physics, School of Medicine, University of Patras, 265 04 Patras, Greece.
Objective:
To estimate organs' absorbed dose from the two-phase CT of parathyroid glands, effective dose (ED) based on three different methods, and compare the dose values with those reported by other published protocols.
Methods:
Volumetric-computed-tomography-dose-index (CTDIvol), dose-length-product (DLP), and the corresponding scan length during each phase of a parathyroid protocol were recorded, for seventy-six patients. One k-factor, and two different k-factors for the neck and chest area were used to estimate the ED from DLP. A Monte Carlo software, VirtualDoseCT, was also used for the estimation of organs' absorbed dose and ED.
Results:
Two-phase parathyroid CT resulted in a mean ED of 3.93 mSv, 4.29 mSv and 4.21 mSv according to the one k-factor, two k-factors, and VirtualDoseCT methods, respectively. The two k-factors method resulted in a slight overestimation of 1.9% in total ED compared to VirtualDoseCT. No statistically significant difference was found in ED values between these methods (Wilcoxon test, p > 0.05), except for female patients in the pre-contrast phase. The organs inside the scanning field of view (SFOV) received the following doses: thymus 23.3 mGy, lungs 11.5 mGy, oesophagus 9.2 mGy, thyroid 6.9 mGy, and breast 6.3 mGy. The ED and organs' dose (OD) values were significantly lower in the pre-contrast than in the arterial phase (Wilcoxon test, p < 0.001). A statistically significant difference was observed between male and female patients for the pre-contrast phase (Mann-Whitney test, p < 0.05), regarding the ED values obtained with the two k-factors method and VirtualDoseCT software.
Conclusions:
The two k-factors method could be applied for the ED estimation in clinical practice, if appropriate software is not available. An extensive range of ED values derived from the literature, mainly depending on the acquisition protocol parameters and the estimation method.
More Related Videos
07:12Establishment of a Simple and Effective Rat Model for Intraoperative Parathyroid Gland Imaging
Published on: August 17, 2022
07:13Two Techniques to Create Hypoparathyroid Mice: Parathyroidectomy Using GFP Glands and Diphtheria-Toxin-Mediated Parathyroid Ablation
Published on: March 14, 2017
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
The Parathyroid Glands
Oxyphil cells, whose functions remain elusive, emerge during late puberty, adding a layer of complexity to the parathyroid gland's intricacies. In contrast, principal parathyroid cells undertake a vital role by...
Biological Effects of Radiation