Related Experiment Video
Updated: Sep 21, 2025

Determining and Controlling External Power Output During Regular Handrim Wheelchair Propulsion
Published on: February 5, 2020
Quality inspection and result analysis of the spirometer calibration cylinder
Zhongping Wu1, Yongyi Peng1, Kuiqing Lin1
1National Center for Respiratory Medicine, State Key Laboratory of Respiratory Disease, National Clinical Research Center for Respiratory Disease, Guangzhou Institute of Respiratory Health, First Affiliated Hospital of Guangzhou Medical University, Yanjiang Road 151, Guangzhou, 510120, People's Republic of China.
Background:
To understand the accuracy of volume calibration syringes used in China and compare the difference between new and old volume calibration syringes, technical testing was performed on volume calibration syringes in clinical lung function instruments.
Materials And Methods:
A standard validator device (Model 1180, Hans Rudolph, USA) was used to perform leak testing and volume accuracy testing for calibration syringes. Sixteen volume calibration syringes from 8 brands (CareFusion in Germany, Vyaire in Germany, Yaeger in Germany, Vitalograph in the United Kingdom, MGC Diagnostics in the United States, U-Breath in Zhejiang, China, Wendi in Ningbo, Zhejiang, and Boya in Ningbo, China) were tested.
Results:
A total of 75% (12/16) of the volume calibration syringes passed the pressure decay leak test, 69% (11/16) of the volume calibration syringes passed the volume accuracy and repeatability test, and 56% (9/16) passed both tests; there was no significant difference in the total passing of the new and old volume calibration syringe quality tests (P > 0.05).
Conclusions:
A standard validator device should be used for both leakage tests and volume accuracy and repeatability tests to ensure the reliability of volume calibration syringes. It is suggested that the quality verification of volume calibration syringes should be regularly conducted to ensure the accuracy of the pulmonary function tests.
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Quality Assurance
Uncertainty in Measurement: Reading Instruments
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...