An Objective and Child-friendly Assessment of Arm Function by Using a 3-D Sensor
Published on: February 12, 2018
Reliability of 3D Depth Motion Sensors for Capturing Upper Body Motions and Assessing the Quality of Wheelchair
Alicia Marie Koontz1,2,3, Ahlad Neti1,2, Cheng-Shiu Chung1,3
1Human Engineering Research Laboratories, VA Pittsburgh Healthcare System, Pittsburgh, PA 15206, USA.
Abstract:
Wheelchair users must use proper technique when performing sitting-pivot-transfers (SPTs) to prevent upper extremity pain and discomfort. Current methods to analyze the quality of SPTs include the TransKinect, a combination of machine learning (ML) models, and the Transfer Assessment Instrument (TAI), to automatically score the quality of a transfer using Microsoft Kinect V2. With the discontinuation of the V2, there is a necessity to determine the compatibility of other commercial sensors. The Intel RealSense D435 and the Microsoft Kinect Azure were compared against the V2 for inter- and intra-sensor reliability. A secondary analysis with the Azure was also performed to analyze its performance with the existing ML models used to predict transfer quality. The intra- and inter-sensor reliability was higher for the Azure and V2 (n = 7; ICC = 0.63 to 0.92) than the RealSense and V2 (n = 30; ICC = 0.13 to 0.7) for four key features. Additionally, the V2 and the Azure both showed high agreement with each other on the ML outcomes but not against a ground truth. Therefore, the ML models may need to be retrained ideally with the Azure, as it was found to be a more reliable and robust sensor for tracking wheelchair transfers in comparison to the V2.