Wearable Devices and Smartphone Inertial Sensors for Static Balance Assessment: A Concurrent Validity Study in Young
Luciana Abrantes Rodrigues1, Enzo Gabriel Rocha Santos2, Patrícia Seixas Alves Santos3
1Núcleo de Medicina Tropical, Universidade Federal do Pará, Belém 66050-540, Brazil.
Abstract:
Falls represent a public health issue around the world and prevention is an important part of the politics of many countries. The standard method of evaluating balance is posturography using a force platform, which has high financial costs. Other instruments, such as portable devices and smartphones, have been evaluated as low-cost alternatives to the screening of balance control. Although smartphones and wearables have different sizes, shapes, and weights, they have been systematically validated for static balance control tasks. Different studies have applied different experimental configurations to validate the inertial measurements obtained by these devices. We aim to evaluate the concurrent validity of a smartphone and a portable device for the evaluation of static balance control in the same group of participants. Twenty-six healthy and young subjects comprised the sample. The validity for static balance control evaluation of built-in accelerometers inside portable smartphone and wearable devices was tested considering force platform recordings as a gold standard for comparisons. A linear correlation (r) between the quantitative variables obtained from the inertial sensors and the force platform was used as an indicator of the concurrent validity. Reliability of the measures was calculated using Intraclass correlation in a subsample (n = 14). Smartphones had 11 out of 12 variables with significant moderate to very high correlation (r > 0.5, p < 0.05) with force platform variables in open eyes, closed eyes, and unipedal conditions, while wearable devices had 8 out of 12 variables with moderate to very high correlation (r > 0.5, p < 0.05) with force platform variables under the same task conditions. Significant reliabilities were found in closed eye conditions for smartphones and wearables. The smartphone and wearable devices had concurrent validity for the static balance evaluation and the smartphone had better validity results than the wearables for the static balance evaluation.
More Related Videos
06:49Evaluation of a Smartphone-based Human Activity Recognition System in a Daily Living Environment
Published on: December 11, 2015
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025