Related Experiment Video
Updated: Sep 3, 2025

Building Langmuir Probes and Emissive Probes for Plasma Potential Measurements in Low Pressure, Low Temperature Plasmas
Published on: May 25, 2021
Development of the Measurement of Lateral Electron Density (MOLE) Probe Applicable to Low-Pressure Plasma Diagnostics
Si-Jun Kim1, Sang-Ho Lee1,2, Ye-Bin You1
1Applied Physics Lab for PLasma Engineering (APPLE), Department of Physics, Chungnam National University, Daejeon 34134, South Korea.
Abstract:
As the importance of measuring electron density has become more significant in the material fabrication industry, various related plasma monitoring tools have been introduced. In this paper, the development of a microwave probe, called the measurement of lateral electron density (MOLE) probe, is reported. The basic properties of the MOLE probe are analyzed via three-dimensional electromagnetic wave simulation, with simulation results showing that the probe estimates electron density by measuring the surface wave resonance frequency from the reflection microwave frequency spectrum (S11). Furthermore, an experimental demonstration on a chamber wall measuring lateral electron density is conducted by comparing the developed probe with the cutoff probe, a precise electron density measurement tool. Based on both simulation and experiment results, the MOLE probe is shown to be a useful instrument to monitor lateral electron density.
Related Concept Videos
Atomic Emission Spectroscopy: Lab
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview