Smartwatch-Based Blood Pressure Measurement Demonstrates Insufficient Accuracy
Maarten Falter1,2,3, Martijn Scherrenberg1,2,4, Karen Driesen1
1Faculty of Medicine and Life Sciences, Hasselt University, Hasselt, Belgium.
Frontiers in Cardiovascular Medicine
|July 28, 2022
Summary
This study found that the Samsung Galaxy Watch Active 2 cuffless blood pressure monitor shows significant bias, overestimating low readings and underestimating high ones. The device is not yet clinically ready for accurate blood pressure measurement.
Area of Science:
- Biomedical Engineering
- Cardiovascular Technology
- Wearable Health Devices
Background:
- Novel cuffless blood pressure (BP) measuring smartwatches are emerging with regulatory approval.
- These devices often lack sufficient clinical validation.
- This study evaluates a CE-cleared smartwatch for BP measurement in diverse patient groups.
Purpose of the Study:
- To investigate the accuracy and reliability of a specific CE-cleared smartwatch for cuffless BP measurement.
- To assess BP measurement bias and precision in normotensive and hypertensive individuals.
- To compare BP variability between smartwatch and traditional ambulatory monitoring.
Main Methods:
- Simultaneous BP measurements using a Samsung Galaxy Watch Active 2 and 24-h ambulatory BP monitoring (ABPM).
- Calibration of the smartwatch followed by 24-h monitoring.
- Accuracy assessment via sensitivity, specificity, predictive values, ROC curves, Bland-Altman, and Taffé methods.
- BP variability analysis using standard deviation, coefficient of variation, and average real variability.
Main Results:
- Forty patients participated; proportional bias was observed in systolic and diastolic BP measurements.
- Low systolic BPs were overestimated, high BPs underestimated; diastolic BPs were consistently overestimated.
- Sensitivity and specificity for hypertension detection were 83% and 41%; AUCs were 0.78 (systolic) and 0.93 (diastolic).
- BP variability was systematically higher with ABPM than the smartwatch.
Conclusions:
- The Samsung Galaxy Watch Active 2 exhibits systematic bias, rendering it unsuitable for clinical use currently.
- Existing standards for non-invasive sphygmomanometers are not met and may not apply to cuffless devices.
- New validation standards are urgently needed for wearable BP monitors, alongside further research into non-invasive, high-frequency BP monitoring.
Related Concept Videos
Errors occurring during blood pressure monitoring
959
Blood pressure monitoring is a crucial clinical procedure in diagnosing and managing various cardiovascular conditions. Despite its significance, the accuracy of blood pressure measurements can be compromised by multiple factors, potentially leading to either falsely high or low readings. These inaccuracies are critical as they can significantly impact patient care. So, it is vital to understand these challenges deeply and adopt strategic approaches to minimize errors.
Several factors...
Several factors...
959
Assessment of blood pressure in brachial artery(two-step method)
875
Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
875
Special considerations while measuring blood pressure
813
When assessing blood pressure (BP), healthcare professionals must consider various factors and potential unexpected outcomes to ensure accurate readings and provide proper patient care. Adhering to these guidelines is essential to achieving the most reliable results.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
813
Measurement of Blood Pressure
1.3K
Assessing blood pressure is a standard procedure executed in virtually all medical environments. The method utilized today was established over a hundred years ago by an innovative Russian doctor, Dr. Nikolai Korotkoff. The soft ticking noise, known as Korotkoff sounds, heard while taking blood pressure readings results from turbulent blood flow within the vessels. The apparatus required for this procedure includes a sphygmomanometer, a blood pressure cuff attached to a gauge, and a...
1.3K
Equipments Used To Measure Blood Pressure
1.3K
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
1.3K
Sites for measruring blood pressure
2.3K
Blood pressure measurement is a fundamental clinical procedure, providing crucial data for assessing cardiovascular health. Among the various sites for this measurement, the brachial and popliteal arteries are predominantly utilized due to their accessibility and the reliability of their readings. This lesson delves into the anatomical significance, methodology, and considerations of measuring blood pressure at these locations.
The Brachial Artery: Primary Site for Blood Pressure Measurement
The Brachial Artery: Primary Site for Blood Pressure Measurement
2.3K