Effectuality study of a 3D motion correction algorithm in C-arm CTs of severely impaired image quality during
Lena S Becker1, Cornelia L A Dewald2, Christian von Falck2
1Department of Diagnostic and Interventional Radiology, Hannover Medical School, OE8220 Carl-Neuberg-Str. 1, 30625, Hannover, Germany. becker.lena@mh-hannover.de.
Background:
To evaluate effectivity of a 3D-motion correction algorithm in C-Arm CTs (CACT) with limited image quality (IQ) during transarterial chemoembolization (TACE).
Methods:
From 1/2015-5/2021, 644 CACTs were performed in patients during TACE. Of these, 27 CACTs in 26 patients (18 m, 8f; 69.7 years ± 10.7 SD) of limited IQ were included. Post-processing of the original raw-data sets (CACTOrg) included application of a 3D-motion correction algorithm and bone segmentation (CACTMC_no_bone). Four radiologists (R1-4) compared the images by choosing their preferred dataset and recommending repeat acquisition in case of severe IQ-impairment. R1,2 performed additional grading of intrahepatic vessel visualization, presence/extent of movement artifacts, and overall IQ.
Results:
R1,2 demonstrated excellent interobserver agreement for overall IQ (ICC 0.79,p < 0.01) and the five-point vessel visualization scale before and after post-processing of the datasets (ICC 0.78,p < 0.01). Post-processing caused significant improvement, with overall IQ improving from 2.63 (CACTOrg) to 1.39 (CACTMC_no_bone;p < 0.01) and a decrease in the mean distance of identifiable, subcapsular vessels to the liver capsule by 4 mm (p < 0.01). This proved especially true for datasets with low parenchymal and high hepatic artery contrast. A good interobserver agreement (ICC = 0.73) was recorded concerning the presence of motion artifacts, with significantly less discernible motion after post-processing (CACTOrg:1.31 ± 1.67, CACTMC_no_bone:1.00 ± 1.34, p < 0.01). Of the 27 datasets, ≥ 23 CACTMC_no_bone were preferred, with identical datasets chosen by the readers to show benefit from the algorithm.
Conclusion:
Application of a 3D-motion correction algorithm significantly improved IQ in diagnostically limited CACTs during TACE, with the potential to decrease repeat acquisitions.
Related Concept Videos
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...