Related Experiment Video
Updated: Aug 30, 2025

09:11
Y-90 Radioembolization and PD-1 Inhibitor as Neoadjuvant Treatment in Hepatocellular Carcinoma
Published on: May 24, 2024
577
Posttreatment Exposure Rates for 90Y-Microsphere Patients: A Comparison of Products
Steven Blum1, Eugenio Silvestrini2,3, Jonathan Weinstein4
1Hofstra University, Hempstead, New York.
Journal of Nuclear Medicine Technology
|August 30, 2022
Summary
Radioactive yttrium-90 (90Y) microspheres are increasingly used for liver cancer. A study found no significant difference in radiation exposure rates between TheraSphere and SIR-Spheres, meaning no special patient considerations are needed.
Area of Science:
- Nuclear Medicine
- Oncology
- Radiology
Background:
- 90Y-microsphere therapy has grown significantly for liver malignancies over the past 30 years.
- FDA approval of Sirtex SIR-Spheres and Boston Scientific TheraSphere has driven this increase.
- Institutional experience with both 90Y-microsphere products necessitated comparative analysis.
Purpose of the Study:
- To compare radiation exposure rates between TheraSphere and SIR-Spheres.
- To determine if product-specific considerations are necessary for patient safety and radiation protection.
- To ensure adherence to as-low-as-reasonably-achievable (ALARA) principles.
Main Methods:
- An independent-samples t-test (α = 0.05) was used for analysis.
- Radiation exposure rates were measured in multiple patient regions pre- and post-implantation.
- The study included 50 patients: 25 treated with TheraSphere and 25 with SIR-Spheres.
Main Results:
- No statistically significant differences in radiation exposure rates were observed between the two 90Y-microsphere products.
- Both TheraSphere and SIR-Spheres demonstrated similar radiation profiles.
- The findings support the application of uniform safety protocols for both treatments.
Conclusions:
- TheraSphere and SIR-Spheres exhibit comparable radiation exposure rates in patients undergoing treatment for liver malignancies.
- No unique procedural considerations are required for one 90Y-microsphere product over the other.
- ALARA principles can be effectively maintained with either treatment modality without product-specific adjustments.