Flow Cytometric Analysis of Biomarkers for Detecting Human Sperm Functional Defects
Published on: April 21, 2022
2.6K
Staining-free, Automated Sperm Analysis for In Vitro Fertilization Lab Use
Changsheng Dai1, Guanqiao Shan1, Sahar Jahangiri2
1Department of Mechanical and Industrial Engineering, University of Toronto, Toronto, Ontario, Canada.
The Journal of Urology
|September 13, 2022
Summary
Automated sperm analysis offers a reliable and reproducible method for assessing sperm concentration, motility, and morphology. This system is suitable for routine use in in vitro fertilization labs, improving quantitative analysis of live sperm samples.
Area of Science:
- Reproductive biology andrology
- In vitro fertilization (IVF) laboratory techniques
- Medical imaging and artificial intelligence
Background:
- Traditional sperm analysis in IVF labs relies on subjective manual observation, leading to inconsistencies.
- Current computer-aided sperm analysis systems are typically for andrology labs and require sperm staining for morphology.
- A need exists for objective, automated sperm analysis in IVF settings without staining.
Purpose of the Study:
- To develop and evaluate an automated system for simultaneous measurement of sperm concentration, motility, and morphology.
- To assess the system's performance without the requirement of sperm staining.
- To determine the reproducibility and reliability of the automated sperm analysis system.
Main Methods:
- Utilized robust sperm tracking for automated measurement of sperm concentration and motility.
- Employed low contrast image segmentation for automated morphology assessment of live sperm.
- Evaluated measurement reproducibility using intraclass correlation coefficients and reliability via Passing and Bablok regression and Bland-Altman analysis.
Main Results:
- Automated sperm measurements demonstrated high reproducibility with intraclass correlation coefficients exceeding 0.97.
- Automated measurements were found to be interchangeable with manual benchmarking using zoomed-in images.
- The system showed high reproducibility and reliability across both semen and washed samples, including progressive and nonprogressive motility.
Conclusions:
- The developed automated sperm analysis system exhibits high reproducibility and reliability.
- This system is designed for routine application in in vitro fertilization laboratories.
- It enables quantitative sperm analysis on live samples, overcoming limitations of manual methods.