A Magnetic Resonance Imaging Protocol for Stroke Onset Time Estimation in Permanent Cerebral Ischemia
Published on: September 16, 2017
Synthetic MRI with Magnetic Resonance Spin TomogrAphy in Time-Domain (MR-STAT): Results from a Prospective
Jordi P D Kleinloog1,2, Stefano Mandija1,2, Federico D'Agata3
1Computational Imaging Group for MR Therapy and Diagnostics, Center for Image Sciences, University Medical Center Utrecht, Utrecht, The Netherlands.
Background:
Magnetic Resonance Spin TomogrAphy in Time-domain (MR-STAT) can reconstruct whole-brain multi-parametric quantitative maps (eg, T1 , T2 ) from a 5-minute MR acquisition. These quantitative maps can be leveraged for synthetization of clinical image contrasts.
Purpose:
The objective was to assess image quality and overall diagnostic accuracy of synthetic MR-STAT contrasts compared to conventional contrast-weighted images.
Study Type:
Prospective cross-sectional clinical trial.
Population:
Fifty participants with a median age of 45 years (range: 21-79 years) consisting of 10 healthy participants and 40 patients with neurological diseases (brain tumor, epilepsy, multiple sclerosis or stroke).
Field Strength/Sequence:
3T/Conventional contrast-weighted imaging (T1 /T2 weighted, proton density [PD] weighted, and fluid-attenuated inversion recovery [FLAIR]) and a MR-STAT acquisition (2D Cartesian spoiled gradient echo with varying flip angle preceded by a non-selective inversion pulse).
Assessment:
Quantitative T1 , T2 , and PD maps were computed from the MR-STAT acquisition, from which synthetic contrasts were generated. Three neuroradiologists blinded for image type and disease randomly and independently evaluated synthetic and conventional datasets for image quality and diagnostic accuracy, which was assessed by comparison with the clinically confirmed diagnosis.
Statistical Tests:
Image quality and consequent acceptability for diagnostic use was assessed with a McNemar's test (one-sided α = 0.025). Wilcoxon signed rank test with a one-sided α = 0.025 and a margin of Δ = 0.5 on the 5-level Likert scale was used to assess non-inferiority.
Results:
All data sets were similar in acceptability for diagnostic use (≥3 Likert-scale) between techniques (T1 w:P = 0.105, PDw:P = 1.000, FLAIR:P = 0.564). However, only the synthetic MR-STAT T2 weighted images were significantly non-inferior to their conventional counterpart; all other synthetic datasets were inferior (T1 w:P = 0.260, PDw:P = 1.000, FLAIR:P = 1.000). Moreover, true positive/negative rates were similar between techniques (conventional: 88%, MR-STAT: 84%).
Data Conclusion:
MR-STAT is a quantitative technique that may provide radiologists with clinically useful synthetic contrast images within substantially reduced scan time.
Evidence Level:
1 Technical Efficacy: Stage 2.
More Related Videos
13:56The Use of Magnetic Resonance Spectroscopy as a Tool for the Measurement of Bi-hemispheric Transcranial Electric Stimulation Effects on Primary Motor Cortex Metabolism
Published on: November 19, 2014
09:30Quantitative Magnetic Resonance Imaging of Skeletal Muscle Disease
Published on: December 18, 2016
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...