Wideband absorbance for the assessment of pressure equalizing tubes patency in children
Eliene Silva Araújo1, Lilian Cassia Bornia Jacob2, Maria Taiany Duarte de Oliveira3
1Department of Speech-Language Pathology and Audiology, Federal University of Rio Grande do Norte, Natal, Rio Grande do Norte, Brazil.
Objectives:
To analyze the feasibility of using wideband absorbance to verify the patency of pressure equalizing tubes (PETs) in clinical practice and to present the response pattern of this measure for ears with patent PET.
Methods:
This observational case-control type study evaluated 48 ears of 30 children with severe or profound hearing loss, aged 10-44 months, of both sexes. The subjects were subdivided into two groups: 24 ears with Sheppard type PET (experimental group - EG) and 24 ears with normal middle ear (control group - CG), paired with the EG, according to age, sex, and ear evaluated. To obtain the wideband absorbance, a Middle-Ear Power Analyzer, version 5.0 (Mimosa Acoustics), was used, and absorbance values for pure tone and chirp stimuli were analyzed.
Results:
There was no influence of ear (right or left) on the measurements obtained. The EG showed higher absorbance values at low frequencies. Although the two stimuli made it possible to identify the difference in acoustic transfer function between the groups studied, compared to pure tone, the chirp stimulus allowed identification of differences in a higher number of frequencies.
Conclusions:
Ears with a patent PET present an acoustic transfer pattern that differs from that obtained for normal middle ears, with a higher absorbance at low frequencies. Both pure tone and chirp stimuli can be used to identify such differences, nevertheless, the use of chirp stimulus is recommended, since it allows differentiation over a wider frequency range.
Related Concept Videos
Pre-Procedural Guidelines for Assessing Blood Pressure
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: