Photon-counting detector CT improves quality of arterial phase abdominal scans: A head-to-head comparison with
D Graafen1, L Müller1, M Halfmann2
1Diagnostic and Interventional Radiology, University Medical Center of the Johannes Gutenberg-University Mainz, Mainz, Germany.
Purpose:
Photon-counting detector (PCD)-CT is expected to have a substantial impact on oncologic abdominal imaging. We compared subjective and objective image quality between PCD-CT and conventional energy-integrating detector (EID-)CT arterial phase abdominal scans.
Methods:
This study included 84 patients undergoing both types of abdominal CT. EID-CT scans were acquired with a tube voltage of 100 kVp. With PCD-CT, acquired with 120-kVp, we reconstructed polychromatic T3D images and virtual monoenergetic images (VMIs) in 10-keV intervals from 40 to 90 keV. Quantitative image analysis included noise and contrast-to-noise ratio (CNR) of hepatic vessels, kidney cortex, and hypervascular liver lesions to liver parenchyma. Three raters used a 5-point Likert scale for qualitative image analysis of image noise and contrast, lesion conspicuity, and overall image quality. Radiation dose exposure (CT dose index) was compared between the two CT types.
Results:
Mean CT dose index and effective dose were respectively 18 % and 26 % lower with PCD-CT versus EID-CT. Compared with EID-CT, CNRs of kidney cortex and vessel to liver parenchyma were significantly higher in PCD-CT VMIs at energies ≤ 60 keV and in polychromatic T3D images (p < 0.004). Overall image quality of PCD-CT VMIs at 50 and 60 keV was rated as significantly better (p < 0.01) than the EID-CT images (inter-reader agreement alpha = 0.80). Lesion conspicuity was significantly better in low-keV VMIs (p < 0.03) and worse in > 70-keV VMIs.
Conclusions:
With low-keV VMI, PCD-CT yields significantly improved objective and subjective quality of arterial phase oncological imaging compared with EID-CT. This advantage may translate into higher diagnostic confidence and lower radiation dose protocols.
More Related Videos
08:41Novel In Vivo Micro-Computed Tomography Imaging Techniques for Assessing the Progression of Non-Alcoholic Fatty Liver Disease
Published on: March 24, 2023
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET