Lead switching to resolve undersensing of ventricular tachycardia by a cardiac resynchronization therapy
Shohei Kataoka1, Daigo Yagishita1,2, Kyoichiro Yazaki1
1Department of Cardiology, Tokyo Women's Medical University, Tokyo, Japan.
Abstract:
A 63-year-old man was admitted to the hospital due to ventricular tachycardia (VT) undersensing, caused by decreased R-wave amplitude in a cardiac resynchronization therapy defibrillator. The R-wave amplitude of VT sensed by the left ventricular (LV) lead was markedly higher than that by the right ventricular (RV) lead; therefore, we reconnected the IS-1 RV lead to the LV IS-1 port and the IS-1 LV lead to the RV IS-1 port to resolve this issue. After discharge, it was confirmed that VT was successfully terminated by the second sequence of intrinsic ATP (iATP, Medtronic, Minneapolis, MN, USA) from the LV lead.
More Related Videos
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Cardiomyopathy II: Dilated Cardiomyopathy
Cardiopulmonary Resuscitation III: AED Use
Dysrhythmias VI: Management of Dysrhythmias
Cardiomyopathy V: Interprofessional Care
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias I: Introduction