Experimental study on the accuracy of non-invasive temperature measurement by magnetic resonance in microwave
Rui Guo1, Xianjian Wang2, Jie Liu3
1Department of Interventional Radiology, First Affiliated Hospital of Fujian Medical University, Fuzhou, China.
Aims:
This study aimed to explore the accuracy of non-invasive temperature measurement based on proton resonance frequency (PRF) phase subtraction in microwave ablation (MWA).
Methods And Material:
The signal change of the agar phantom during the ablation process was monitored by the gradient echo sequence under 1.5 T superconducting magnetic resonance imagining (MRI), and the temperature change was converted by the phase subtraction method of the PRF, which was compared with the temperature measured using an optical fiber.
Statistical Analysis Used:
SPSS software version 22.0 was used for data processing, and the independent sample t-test was used for comparative analysis. P < 0.05 indicated statistical significance.
Results:
The maximum error between the MRI temperature measurement and the standard value was 3.61°C, whereas the minimum and average errors were 0.01°C and 1.19°C ± 0.78°C, respectively.
Conclusions:
The temperature measurement technology, which is based on the PRF phase method, has good accuracy in MRI-guided MWA.
More Related Videos
09:01Magnetic-, Acoustic-, and Optical-Triple-Responsive Microbubbles for Magnetic Hyperthermia and Pothotothermal Combination Cancer Therapy
Published on: May 22, 2020
13:41Magnetic Resonance-Guided High Intensity Focused Ultrasound Generated Hyperthermia: A Feasible Treatment Method in a Murine Rhabdomyosarcoma Model
Published on: January 13, 2023
Related Concept Videos
Magnetic Resonance Imaging
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...