The Ames Clinitek 200/Multistix 9 urinalysis method compared with manual and microscopic methods.
Clinical Chemistry
|September 1, 1987
Summary
The Ames Clinitek 200/Multistix 9 method showed good agreement for urine protein, glucose, and ketones. However, it was less sensitive for leukocyte esterase, nitrite, and blood, and inaccurate for pH measurement.
Area of Science:
- Clinical Chemistry
- Urinalysis
Background:
- Semi-automated urinalysis methods offer potential for efficiency.
- Standard urinalysis methods provide established benchmarks for comparison.
Purpose of the Study:
- To evaluate the performance of the Ames Clinitek 200/Multistix 9 system against standard methods for key urine analytes.
- To assess the accuracy and sensitivity of the semi-automated method for detecting protein, glucose, ketones, leukocyte esterase, blood, nitrite, and pH.
Main Methods:
- 315 urine specimens were analyzed using the Ames Clinitek 200/Multistix 9 and standard reference methods.
- Analytes measured included protein, glucose, ketones, leukocyte esterase, blood, nitrite, and pH.
Main Results:
- Excellent agreement was observed between the Clinitek and standard methods for protein, glucose, and ketones.
- The Clinitek method demonstrated lower sensitivity for detecting leukocyte esterase, nitrite, and blood compared to microscopic examination.
- The Clinitek pH method yielded inaccurate readings, showing falsely low or high values outside the 6.5 threshold.
Conclusions:
- The Ames Clinitek 200/Multistix 9 system is reliable for quantifying urine protein, glucose, and ketones.
- The semi-automated method's sensitivity is insufficient for detecting pyuria, bacteriuria, and hematuria compared to traditional microscopic analysis.
- The Clinitek pH measurement requires caution due to observed inaccuracies.