Diagnostic discrimination of a novel high-sensitivity cardiac troponin I assay and derivation/validation of an

Luca Koechlin1, Jasper Boeddinghaus2, Pedro Lopez-Ayala3

  • 1Cardiovascular Research Institute Basel (CRIB) and Department of Cardiology, University Hospital Basel, University of Basel, Basel, Basel, Switzerland; Department of Cardiac Surgery, University Hospital Basel, University of Basel, Basel, Basel, Switzerland; GREAT network, Basel, Basel, Switzerland.

American Heart Journal
|October 15, 2022
PubMed

Insights

The Dimension EXL LOCI High-Sensitivity Troponin I (hs-cTnI-EXL) assay demonstrates diagnostic utility comparable to other leading assays for acute myocardial infarction (MI). An assay-specific algorithm effectively rules out and rules in MI, improving emergency department diagnostics.

Area of Science:

  • Cardiology
  • Clinical Diagnostics
  • Biomarker Assays

Background:

  • Assessing the diagnostic performance of novel high-sensitivity troponin assays is crucial for timely acute myocardial infarction (MI) diagnosis.
  • The Dimension EXL LOCI High-Sensitivity Troponin I (hs-cTnI-EXL) assay's utility required evaluation against established methods.

Purpose of the Study:

  • To evaluate the diagnostic accuracy of the hs-cTnI-EXL assay for acute myocardial infarction (MI).
  • To compare the discriminatory performance of hs-cTnI-EXL with hs-cTnI-Architect and hs-cTnT-Elecsys assays.
  • To derive and validate a hs-cTnI-EXL-specific 0/1-hour algorithm for MI rule-out and rule-in.

Main Methods:

  • Multicenter study involving patients with chest discomfort in the emergency department.
  • Central adjudication of diagnoses by cardiologists, including serial measurements of hs-cTnI-Architect and hs-cTnT-Elecsys.
  • Receiver operating characteristic (ROC) curve analysis to assess discriminatory performance.

Main Results:

  • The hs-cTnI-EXL assay showed an area under the ROC curve of 0.94, comparable to hs-cTnI-Architect (0.95) and hs-cTnT-Elecsys (0.93).
  • A derived 0/1-hour algorithm in the validation cohort achieved 97.8% sensitivity and 98.3% specificity for ruling out and ruling in MI, respectively.
  • The algorithm safely ruled out MI in 56% of patients with a negative predictive value of 99.5%.

Conclusions:

  • The hs-cTnI-EXL assay exhibits diagnostic performance comparable to existing high-sensitivity troponin assays.
  • An assay-specific 0/1-hour algorithm provides a safe and accurate method for ruling out and ruling in acute MI.
  • This assay and algorithm can aid in efficient and accurate MI diagnosis in emergency settings.