Low partial pressure of oxygen causes significant and unrecognized under-recovery of glucose on blood gas analyzers
Cody Wayne Lewis1, Yury Butorin2, Lawrence de Koning1
1Alberta Precision Laboratories, Calgary, Alberta, Canada; Department of Pathology and Laboratory Medicine, University of Calgary Cumming School of Medicine, Calgary, Alberta, Canada.
Background:
Blood gas analyzers employing glucose-oxidase biosensors under-recover glucose when pO2 is low. The manufacturer of the GEM®Premier™ series of analyzers introduced an algorithm to detect specimens at risk of low pO2 interference. We investigated the reliability of this algorithm.
Methods:
Whole blood specimens were tested by GEM®Premier™ 4000 (GEM 4000) and 5000 (GEM 5000). Specimens with an incalculable ("incalc") error code for glucose result or that had a glucose ≥ 20 mmol/L were retested on a second analyzer of the same type within 5 min over the course of 30 months in 5 hospitals in Calgary, Alberta. Discordant retests were defined as either: 1) paired numeric results with a difference >10 %, or 2) an "incalc" code that yielded a numeric result upon retesting. Glucose recovery in relation to pO2 level was assessed by comparing specimens experimentally depleted of pO2 between GEM 5000 and a laboratory analyzer (Siemens Vista®).
Results:
Of 1,776 glucose tests repeated on the GEM 5000 or 1,544 on GEM 4000, 10% were discordant. GEM 5000 produced twice as many discordant numeric retests versus the GEM 4000 [5.9% (98/1,651) vs 2.7% (38/1,391)]. The majority of "incalc" error codes repeated with a numeric glucose result on both GEM analyzers [(79.7% (122/153) vs 75.2% (94/125)]. Among specimens experimentally depleted of pO2, the GEM 5000 under-recovered glucose by up to 30% compared to the Siemens Vista and were not flagged by an "incalc" code.
Conclusions:
The algorithm in the GEM®PremierTM series of analyzers that flags specimens at risk for glucose under-recovery due to low pO2 does not reliably detect specimens at risk for glucose under-recovery.
More Related Videos
05:21Operation of Laboratory Photobioreactors with Online Growth Measurements and Customizable Light Regimes
Published on: October 28, 2021
12:32High-resolution Respirometry to Measure Mitochondrial Function of Intact Beta Cells in the Presence of Natural Compounds
Published on: January 23, 2018
Related Concept Videos
Oxygen Transport in the Blood
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Gas Exchange and Transport
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include: