Clinical comparison of two automated audiometry procedures
Hui Liu1,2, Bingqing Du1,2, Bo Liu1,2
1Department of Otolaryngology, Head and Neck Surgery, Beijing Tongren Hospital, Capital Medical University, Beijing, China.
Automated pure-tone audiometry offers comparable hearing thresholds to manual tests. A shortened ascending protocol significantly reduces testing time compared to the standard ascending method.
Area of Science:
- Audiology
- Hearing Science
- Medical Technology
Background:
- Automated pure-tone audiometry (APTA) shows potential for accurate hearing threshold estimation.
- Previous studies indicate varying correlations between APTA and manual audiometry across different frequencies.
- Differences in test procedures and equipment may influence APTA results.
Purpose of the Study:
- To compare hearing threshold estimates obtained from two automated audiometry protocols with manual audiometry.
- To evaluate testing time, number of responses, and subject preference for automated versus manual audiometry.
- To identify factors contributing to discrepancies in hearing threshold measurements.
Main Methods:
- One hundred subjects (18-36 years) underwent air-conduction pure-tone audiometry.
- Two automated protocols (ascending and shortened ascending) were compared against manual audiometry.
- Measurements included testing time, total responses, and subject preference.
Main Results:
- A significant difference in threshold distribution was observed at 250 Hz between automated methods.
- The shortened ascending protocol significantly reduced testing time (5.8 min) compared to the ascending protocol (9.8 min).
- Subject preference did not differ significantly between automated and manual procedures.
Conclusions:
- The shortened ascending protocol is more time-efficient for automated audiometry.
- Differences in automated thresholds at 250 Hz stem from distinct test procedures.
- Discrepancies at 8,000 Hz may relate to transducer types and calibration variations.