Effect of Native Aortic Leaflet Geometry Modification on Transcatheter Aortic Valve Neo-sinus and Aortic Sinus Flow:
Beatrice E Ncho1, Keshav Kohli2, Vahid Sadri2
1School of Chemical and Biomolecular Engineering, Georgia Institute of Technology, Atlanta, GA, USA.
Cardiovascular Engineering and Technology
|November 10, 2022
Summary
Modifying aortic leaflet geometry after transcatheter aortic valve replacement (TAVR) can reduce blood flow stagnation. Leaflet splay of at least 10 mm may decrease thrombosis risk.
Area of Science:
- Cardiovascular research
- Biomedical engineering
- Medical device innovation
Background:
- Leaflet thrombosis is a serious complication following transcatheter aortic valve replacement (TAVR).
- Blood flow stagnation in the neo-sinus and aortic sinus regions is linked to increased thrombus severity.
- Native aortic leaflet modification is a potential strategy to improve hemodynamics but lacks extensive data.
Purpose of the Study:
- To evaluate the impact of aortic leaflet geometry modification on flow stagnation in the neo-sinus and aortic sinus regions after simulated TAVR.
- To investigate how different leaflet splay geometries affect fluid dynamics in critical post-TAVR areas.
Main Methods:
- Particle image velocimetry was used to measure flow dynamics under varying cardiac outputs (5 LPM and 2.5 LPM).
- Simulations included an intact leaflet (control) and three modified leaflet geometries with increasing splay distances (5 mm, 10 mm, 20 mm).
Main Results:
- Leaflet geometry modification improved flow communication between neo-sinus and aortic sinus regions compared to the control.
- Flow stagnation in the aortic sinus was reduced by over 64% (5 LPM) and 36% (2.5 LPM) across all modified geometries.
- Neo-sinus flow stagnation reduction was observed only with medium (10 mm) and wide (20 mm) splay geometries.
Conclusions:
- Aortic leaflet geometry modification, specifically with at least 10 mm of leaflet splay, can significantly reduce flow stasis.
- These modifications show potential for decreasing the risk of valve thrombosis after TAVR.