Investigating the Feasibility of Using DenseNet to Improve Coronary Calcification Detection in CT
Kuei-Yuan Hou1, Ching-Ching Yang1
1Department of Radiology, Cathay General Hospital, Taipei, Taiwan, ROC (K.Y.H); Department of Medical Imaging and Radiological Sciences, Kaohsiung Medical University, No.100, Shin-Chuan 1st Road, Sanmin Dist., Kaohsiung, Taiwan, 80708, ROC (C.C.Y.); Department of Medical Research, Kaohsiung Medical University Chung-Ho Memorial Hospital, Kaohsiung, Taiwan, ROC (C.C.Y.); Department of Biomedical Imaging and Radiological Sciences, National Yang Ming Chiao Tung University, Taipei, Taiwan, ROC (K.Y.H).
Objectives:
Interscan reproducibility of coronary artery calcium (CAC) scoring can be improved by using a smaller slice thickness but at the cost of higher image noise. This study aimed to investigate the feasibility of using densely connected convolutional network (DenseNet) to reduce the image noise in CAC scans reconstructed with slice thickness < 3 mm for improving coronary calcification detection in CT.
Methods:
Phantom data acquired with QRM and CIRS phantoms were used for model training and testing, where the DenseNet model adopted in this work was a convolutional neural network (CNN) designed for super resolution recovery. After phantom study, the proposed method was evaluated in terms of its ability to improve calcification detection using patient data. The CNN input images (IMGinput) were CAC scans reconstructed with 0.5-, 1.0- and 1.5-mm slice thickness, while CNN label images were CAC scans reconstructed with 3-mm slice thickness (IMG3mm). Region of interest (ROI) analysis was carried out on IMG3mm, IMGinput and CNN output images (IMGoutput). Two-sample t test was used to compare the difference in Hounsfield Unit (HU) values within ROI between IMG3mm and IMGoutput.
Results:
For the calcifications in QRM phantoms, no statistically significant difference was found when comparing the HU values of 400- and 800-HA calcifications identified on IMG3mm to those on IMGoutput with slice thickness of 0.5, 1.0 or 1.5 mm. On the other hand, statistically significant difference was found when comparing the HU values of 200-HA calcifications identified on IMG3mm to those on IMGoutput with a slice thickness of 0.5 and 1.0 mm. Meanwhile, no statistically significant difference was found when comparing the HU values of 200-HA calcifications identified on IMG3mm to those on IMGoutput with a slice thickness of 1.5 mm. As for the rod inserts in CIRS phantoms simulating 9 different tissue types in human body, there was no statistically significant difference between IMG3mm and IMGoutput with slice thickness of 1.5 mm, and all the p values were larger than 0.10. With regards to patient study, more calcification pixels were detected on IMGoutput with a slice thickness of 1.5 mm than on IMG3mm, so calcifications were more clear on the denoised images.
Conclusion:
According to our results, the CNN-based denoising method could reduce statistical noise in IMGinput with a slice thickness of 1.5 mm without causing significant texture change or variation in HU values. The proposed method could improve cardiovascular risk prediction by detecting small and soft calcifications that are barely identified on 3-mm slice images used in conventional CAC scans.
More Related Videos
06:57Author Spotlight: Advancing Cardiovascular Imaging - Introducing the Spatially Weighted Calcium Score for Early Disease Detection
Published on: September 22, 2023
08:02Author Spotlight: Enhanced Quantification of Cardiovascular Calcification Progression for Longitudinal Micro PET/CT Studies in Small Research Animals
Published on: November 15, 2024
Related Concept Videos
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...