How Does Ambient Light Affect the Image Quality of Phosphor Plate Digital Radiography? A Quantitative Analysis Using
Matheus Sampaio-Oliveira1, Luiz Eduardo Marinho-Vieira1, Victor Aquino Wanderley1
1Department of Oral Diagnosis, Division of Oral Radiology, Piracicaba Dental School, University of Campinas (UNICAMP), Piracicaba 13414-903, Brazil.
Abstract:
The aim of this study is to quantitatively evaluate the influence of the duration of ambient light exposure on the image quality of digital radiographs obtained with contemporary phosphor plate (PSP)-based systems. Radiographs of an aluminum step-wedge were obtained using VistaScan and Express systems at five X-ray exposure times: 0.10, 0.20, 0.32, 0.40, and 0.50 s; the resulting dose-area products were, respectively, 21.93, 43.87, 70.19, 87.75, and 109.6 mGycm2. Before PSP read-out, half of the sensitive surface of the PSP plates was exposed to ambient light for 5, 10, 30, 60, and 90 s. The effect of light exposure on brightness, contrast, contrast-to-noise ratio (CNR), signal-to-noise ratio (SNR), and image saturation was compared using ANOVA with the Tukey test (α = 0.05). Ambient light exposure increased brightness and contrast and reduced CNR and SNR in PSP-based radiographs of contemporary digital systems. At the longest X-ray exposure times, ambient light exposure reduced the dark saturation (mean gray values ≤ 1) observed in Express. In conclusion, the negative effects of ambient light observed on the image quality of PSP-based radiographs are directly proportional to the duration of exposure. Clinicians should be aware of such harmful effects when handling and scanning PSP plates in bright environments.
More Related Videos
Related Concept Videos
X-ray Imaging
Photoluminescence: Applications
Variables Affecting Phosphorescence and Fluorescence
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Fluorescence and Phosphorescence: Instrumentation
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...