Related Experiment Video
Updated: Aug 17, 2025

Stimulated Stokes and Antistokes Raman Scattering in Microspherical Whispering Gallery Mode Resonators
Published on: April 4, 2016
Sub-ppm gas phase Raman spectroscopy in an anti-resonant hollow core fiber
Abstract:
We demonstrate recent progress in the development of a Raman gas sensor using a single cladding ring anti-resonant hollow core micro-structured optical fiber (HC-ARF) and a low power pump source. The HC-ARF was designed specifically for low attenuation and wide bandwidth in the visible spectral region and provided low loss at both the pump wavelength (532 nm) and Stokes wavelengths up to a Raman shift of 5000 cm-1. A novel selective core pressurization scheme was also implemented to further reduce the confinement loss, improving the Raman signal enhancement by a factor of 1.9 compared to a standard fiber filling scheme. By exploiting longer lengths of fiber, direct detection of both methane and hydrogen at concentrations of 5 and 10 ppm respectively is demonstrated and a noise equivalent limit-of-detection of 0.15 ppm is calculated for methane.
Related Concept Videos
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and...