Disposable laryngoscope intubation to reduce equipment failure in an emergency out of OR setting - a quality control case study
Related Experiment Video
Updated: Aug 14, 2025

Author Spotlight: Advancing Awake Nasotracheal Intubation with Flexible Video Rhino-Laryngoscopes
Published on: August 2, 2024
Colby G Simmons1, Tobias Eckle2, Dustin Rogers3
1Department of Anesthesiology, University of Colorado School of Medicine, 12401 E. 17th Ave Leprino Bldg #734Anschutz Medical Campus, Aurora, CO, USA. Colby.Simmons@cuanschutz.edu.
Background:
Reusable laryngoscopes have been reported to be superior to disposable laryngoscopes with plastic blades during emergent intubations. Surprisingly, at our institution a quality reporting system revealed a high number of equipment failures with reusable laryngoscopes in an emergency out-of-OR (operating room) setting. As recent studies indicated an improved quality of disposable laryngoscopes, we hypothesized that a thoroughly evaluated disposable laryngoscope would result in less equipment failure in an emergency out-of-OR setting.
Methods:
To perform a more standardized and time efficient analysis, four distinct disposable laryngoscope blade/handle configurations were trialed during standard intubations (n = 4 × 30) in the OR by experienced anesthesia providers who completed a 6-question, Likert-scale/open-ended survey for product evaluation. The 'best' disposable blade was implemented in an emergency out-of-OR setting and equipment failure rates were monitored over a 3-year period.
Results:
Different disposable laryngoscopes were equal regarding sturdiness, illumination and airway visualization. The laryngoscope with the highest overall score was significantly higher scored than the laryngoscope with the lowest overall score. All disposable laryngoscopes were more cost effective than the reusable ones, and the top scored laryngoscope demonstrated the highest 5-year cost-saving ($210 K). Implementation of the top scored disposable laryngoscope into an emergency out-of-OR setting reduced the equipment failure incidence from high 20s to 0.
Conclusion:
Disposable laryngoscopes are cost effective and superior to reusable laryngoscopes in an emergency out-of-OR setting. We demonstrate that the implementation of a disposable laryngoscope in the emergency out-of-OR setting resulted in a near elimination of equipment related quality submissions which ultimately enhances patient safety.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...
Tracheostomy Care I: Pre-procedural Steps
Required Equipment
The equipment necessary for tracheostomy care includes: