A novel electron source for a compact x-ray tube for microbeam radiotherapy with very high dose rates
Christoph Matejcek1, Johanna Winter2, Kurt Aulenbacher3
1Technical University of Munich, School of Medicine and Klinikum rechts der Isar, Department of Radiation Oncology, Munich, Germany; Johannes Gutenberg-Universität Mainz, Institut für Kernphysik, Mainz, Germany.
Abstract:
Microbeam radiotherapy (MRT) is a novel concept in radiation oncology with arrays of alternating micrometer-wide high-dose peaks and low-dose valleys. Preclinical experiments have shown a lower normal tissue toxicity for MRT with similar tumor control rates compared to conventional radiotherapy. A promising candidate for the demanded compact radiation source is the line-focus x-ray tube. Here, we present the setup of a prototype for an electron accelerator being able to provide a suitable x-ray beam for the tube. Several beam dynamic calculations and simulations were performed concerning particle tracking, thermal and electrostatic properties of the electron source, resulting in a proper beamline, including the cathode, the pierce electrode (PE) and the focusing magnets. These parts are discussed separately. The simulations showed that a rectangular cathode with a small width of 0.4mm is mandatory. To quickly shut down the electron beam, an additional voltage of -600V must be applied to the PE. Moreover, the electric field inside the vacuum chamber stays below 10MVm-1 to minimize the risk of field emission. The thermal simulation validates a small displacement of 0.1mm of the heated cathode with respect to the PE, which must be considered during manufacturing of the cathode-PE assembly. The simulations lead to an adequate choice of cathode, electrodes and beamline to achieve the required focal spot of 0.05×20mm2 with a beam current of 0.3A and an electron energy of 300keV. With this setup first MRT experiments with high dose rates up to 10Gys-1 can be executed.
Related Concept Videos
Transmission Electron Microscopy
X-ray Imaging
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Overview of Electron Microscopy
Nuclear Transmutation