The exeter cemented stem provides outstanding long-term fixation and bone load at 15 years follow-up: A systematic
Fabio Mancino1, Haley Tornberg2, Christopher W Jones1,3
1Department of Orthopaedics, The Orthopaedic Research Foundation of Western Australia (ORFWA), 418838Fiona Stanley Hospital, Perth, WA, Australia.
Purpose:
Since its introduction in 1988, the double-tapered polished Exeter cemented stem has been widely adopted in primary total hip arthroplasty (THA). Despite the results coming from the arthroplasty registries have proven great survivorship, the aim of this study was to dig deeper and describe the modes of failure of the Exeter stem at 15 years follow-up while reporting the clinical and radiographic outcomes.
Methods:
A search of PubMed, MEDLINE, and Embase was performed using the Preferred Reporting Items for Systematic Review and Meta-Analyses since inception of database to January 2022. A meta-analysis was performed on stem's failure rates and clinical outcomes using random effects models. Publication bias was assessed with funnel plots.
Results:
Overall, ten studies met the inclusion criteria with 2167 hips at mean 14.8 ± 4.1 years follow-up. The meta-effect estimate for revision rate for stem-related reasons was 3.8% (CI 95% 2.1-5.6, p < 0.01). The meta-effect for revision rate for stem aseptic loosening (AL) was 0.22% (CI 95% 0-0.4, p = 0.048) and for periprosthetic fracture was 0.6% (CI95% 0.3-0.9, p < 0.001). The meta effect estimate for Oxford Hip Score (OHS) at final follow-up was 32.4 (moderate; CI 95% 23.2-41.6, p <0.001) with and heterogeneity among the studies of I2 0%. Radiolucent lines were reported in 5.5% of cases, with 1.0% of cases (21 hips) reported to be progressive.
Conclusion:
Current evidence suggests that the Exeter cemented stem not only has proven long-term outstanding reliability with a revision rate of 3.8%, but also incredibly low revision rates for AL (0.22%) and periprosthetic fracture (0.6%). It is suitable for a variety of indications, and the consistent radiological appearances indicate durable fixation and load transmission while being associated with a remarkably low stem-related complication rate.
More Related Videos
04:19Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
08:20A Reliable and Reproducible Critical-Sized Segmental Femoral Defect Model in Rats Stabilized with a Custom External Fixator
Published on: March 24, 2019
Related Concept Videos
Strength of Cement
For compressive strength tests, ASTM C 109-05 standards prescribe a cement-sand mix ratio of 1:2.75 and a water/cement ratio of 0.485 for making 2-inch cubes. These cubes are mixed, cast, and cured in saturated lime water at 23°C until testing. Flexural strength testing, outlined in...
Soundness of Cement
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the...