Breathing-Adapted Imaging Techniques for Rapid 4-Dimensional Lung Tomosynthesis
Related Experiment Video
Updated: Aug 8, 2025

Author Spotlight: Optimized Lung MRI Protocol with Computationally Efficient Reconstruction Methods
Published on: September 6, 2024
Arielle A Uejo1, Michael G Snyder2, Joseph T Rakowski3
1Department of Oncology, Karmanos Cancer Institute, Flint, MI.
Purpose:
This article presents enhancements to a 4-dimensional (4D) lung digital tomosynthesis (DTS) model introduced in a 2018 paper. That model was proposed as an adjunct to 4D computed tomography (4DCT) to improve tumor localization through artifact reduction achieved by imaging the entire lung in all projections, reducing the projection collection time duration for each phase compared with 4DCT, and requiring only a single-breath cycle to capture all phases. This is applicable to SABR treatment planning. Enhancements comprise customized patient 4D-DTS x-ray scanning parameters.
Methods And Materials:
Imaging parameters derived with the 4D-DTS model were arc duration, frames per second, pulse duration, and tube current normalized to single-chest radiographic milliampere-seconds (mA/mAsAEC). Optimized phase-specific DTS projections imaging parameters were derived for volunteer respiration-tracking surrogate waveforms and for sinusoidal waveforms. These parameters are temporally matched to the respiratory surrogate waveform and presented as continuous data plots during a period of 20 seconds. Comparison is made between surrogate excursions during a single-phase CT and 4D-DTS reconstructions.
Results:
4D-DTS imaging techniques were customized to volunteer respiratory waveforms and sinusoidal waveforms. Technique settings at the highest velocity portions of the volunteer waveforms were arc duration 0.066 seconds, frame rate 921 Hz, pulse duration 1.076 ms, and normalized tube current 76.2 s-1. Technique settings at the highest velocity portions of the sinusoidal waveforms were arc duration 0.029 seconds, frame rate 2074 Hz, pulse duration 0.472 ms, and normalized tube current 173.6 s-1. Sinusoidal surrogate excursion distance at the highest velocity portion of the waveform during a CT rotation of 0.5 seconds ranged from 2.68 to 21.09 mm, all greater than the limiting excursion distance chosen in the 4D-DTS model.
Conclusions:
4D-DTS image technique settings can be customized to individual patient breathing patterns so that captured range of motion satisfies an operator-selected value.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...