Aerosol Concentrations During Otolaryngology Procedures in a Negative Pressure Isolation Room
Kachorn Seresirikachorn1,2, Patthrarawalai Sirinara3,4, Napadon Tangjaturonrasme1
1Department of Otolaryngology, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand.
Objective:
To evaluate the role of a negative pressure room with a high-efficiency particulate air (HEPA) filtration system on reducing aerosol exposure in common otolaryngology procedures.
Study Design:
Prospective quantification of aerosol generation.
Settings:
Tertiary care.
Methods:
The particle concentrations were measured at various times during tracheostomy tube changes with tracheostomy suctioning, nasal endoscopy with suctioning, and fiberoptic laryngoscopy (FOL), which included 5 times per procedure in a negative pressure isolation room with a HEPA filter and additional 5 times in a nonpressure-controlled room without a HEPA filter. The particle concentrations were measured from the baseline, during the procedure, and continued until 30 minutes after the procedure ended. The particle concentrations were compared to the baseline concentrations.
Results:
The particle concentration significantly increased from the baseline during tracheostomy tube changes (mean difference [MD] 0.80 × 106 p/m3 , p = .01), tracheostomy suctioning (MD 0.78 × 106 p/m3 , p = .004), at 2 minutes (MD 1.29 × 106 p/m3 , p = .01), and 3 minutes (MD 1.3 × 106 p/m3 , p = .004) after suctioning. There were no significant differences in the mean particle concentrations among various time points during nasal endoscopy with suctioning and FOL neither in isolation nor nonpressure-controlled rooms.
Conclusion:
A negative pressure isolation room with a HEPA filter was revealed to be safe for medical personnel inside and outside the room. Tracheostomy tube change with tracheostomy suctioning required an isolation room because this procedure generated aerosol, while nasal endoscopy with suctioning and FOL did not. Aerosol generated in an isolation room was diminished to the baseline after 4 minutes.
More Related Videos
05:07Development of a Nose-only Inhalation Toxicity Test Chamber That Provides Four Exposure Concentrations of Nano-sized Particles
Published on: March 18, 2019
07:28Dry Powder and Nebulized Aerosol Inhalation of Pharmaceuticals Delivered to Mice Using a Nose-only Exposure System
Published on: April 6, 2017
Related Concept Videos
Transmission-based Precautions II: Airborne and Protective Environment
Airborne precautions:
Use airborne precautions when treating patients known or suspected to have diseases that spread through the air—for example, tuberculosis or measles. These organisms are present in smaller droplets expelled by an infected person and...
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Suctioning the Nasopharyngeal Airway
Equipment Required
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Tracheostomy Suctioning II: Procedure
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...