Significant Interference of Biotin in Thyroid Function Tests Using Beckman Analyzer: How to Identify such
Melissa McBride1, Amitava Dasgupta2
1Department of Clinical Laboratory Sciences, University of Kansas Medical Center, Kansas City, KS, USA.
Objective:
Biotin in elevated concentration interferes with biotin-based immunoassays. We studied biotin interferences in TSH, FT4, FT3, total T4, total T3 and thyroglobulin assays both in vitro and in vivo using Beckman DXI800 analyzer.
Methods:
Two serum pools were prepared from left-over specimens. Then aliquots of each pool (and serum control) were supplemented with various amounts of biotin followed by measuring thyroid function tests again. Three volunteers each took 10 mg biotin supplement. We compared thyroid function tests before and 2 h after taking biotin.
Results:
We observed significant biotin interferences in biotin-based assays (positive interference with FT4, FT3, and total T3 assay but negative interference with thyroglobulin) both in vitro and in vivo but non-biotin-based assays (TSH and total T4) were not affected.
Conclusions:
Elevated FT3 and FT4 in the presence of normal TSH is inconsistent with hyperthyroidism and should be followed up with total T3 and T4 test. Significant discrepancy between total T3 (falsely elevated value due to biotin) and total T4 (not affected as the assay is not biotin based) maybe an indication of biotin interference.