Use of a Negative Pressure Containment Pod Within Ambulance-Workspace During Pandemic Response.
Mirle Pena1, Dylan T Neu2, H Amy Feng2
1Division of Field Studies and Engineering, National Institute for Occupational Safety and Health, Centers for Disease Control and Prevention, Cincinnati, OH 45226; Department of Biomedical Engineering, University of Cincinnati, Cincinnati, OH 45221.
Emergency medical service (EMS) providers face high infection risks. A new containment pod with HEPA filtration significantly reduced airborne particles in ambulances during aerosol-generating procedures, enhancing provider safety.
Area of Science:
- Occupational Health and Safety
- Infectious Disease Control
- Aerosol Science
Background:
- Emergency medical service (EMS) providers have elevated exposure risks to infectious agents compared to the general population.
- Prehospital environments, particularly ambulances, pose significant risks for disease transmission due to limited space and lack of advanced air filtration.
- Aerosol-generating procedures, such as intubation, performed in ambulances increase pathogen exposure for EMS personnel.
Purpose of the Study:
- To evaluate a novel containment plus filtration intervention designed to reduce aerosol concentrations within ambulance patient modules.
- To assess the effectiveness of a HEPA-filtered extraction system in mitigating exposure risks during aerosol-generating procedures in prehospital settings.
Main Methods:
- Aerosol concentration measurements were conducted in an unoccupied research ambulance using a tracer aerosol and optical particle counters (OPCs).
- Three conditions were tested: baseline (no intervention), containment pod with HEPA-1 filtration, and containment pod with HEPA-2 filtration.
- The intervention involved a containment pod equipped with a high efficiency particulate air (HEPA)-filtered extraction system.
Main Results:
- The containment pod with HEPA-filtered extraction demonstrated significant aerosol reduction capabilities.
- This intervention achieved 95% containment of the total generated particle concentration compared to the baseline condition.
- The system provided rapid air cleaning within the containment pod after aerosol generation.
Conclusions:
- The evaluated containment pod with HEPA-filtered extraction is an effective intervention for reducing aerosol concentrations in ambulance patient modules.
- This system can substantially decrease EMS provider exposure to airborne pathogens during aerosol-generating procedures.
- Implementing such interventions can enhance safety for frontline healthcare workers in prehospital environments.
Related Concept Videos
Transmission-based Precautions II: Airborne and Protective Environment
Airborne precautions:
Use airborne precautions when treating patients known or suspected to have diseases that spread through the air—for example, tuberculosis or measles. These organisms are present in smaller droplets expelled by an infected person and...
PPE Use in Healthcare Settings II: Doffing
Transmission-based Precautions I: Contact, Enteric, and Droplets
Contact Precautions:
Contact precautions are the measures taken to prevent the transmission of infectious agents, especially epidemiologically important microorganisms such as MRSA or influenza, primarily transmitted through direct or indirect contact with an...
Cardiopulmonary Resuscitation II: ACLS Airway Management
PPE Use in Healthcare Settings I: Donning
Standard Precaution
Hand hygiene is the most crucial means to prevent the transmission of disease. Employers are legally required to provide their workers with personal protective equipment (PPE) to minimize exposure or contact with...