Related Experiment Video
Updated: Aug 7, 2025

Optimized Management of Endovascular Treatment for Acute Ischemic Stroke
Published on: January 18, 2018
Flat-panel detector CT to assess intracranial hemorrhage immediately following mechanical thrombectomy
Violiza Inoa1,2,3, David Dornbos4, Julie DiNitto5
1Semmes-Murphey Clinic, Memphis, Tennessee, USA.
Background And Purpose:
The risk of symptomatic intracranial hemorrhage (ICH) approaches 5% despite mechanical thrombectomy (MT) efficacy for ischemic stroke secondary to large vessel occlusion. Flat-panel detector CT (FDCT) imaging with Syngo Dyna CT imaging (Siemens Medical Solutions, Malvern, PA) can be used immediately following MT to detect ICH.
Purpose:
To evaluate the accuracy and reliability of FDCT imaging with Dyna CT compared to conventional post-MT CT and MRI.
Methods:
Head FDCT (20 second, 70 kV) was performed immediately following MT on 26 consecutive patients; postprocedural CT or MRI was obtained ∼24 hours later. Two blinded, independent neuroradiologists evaluated all imaging, identifying ICH, stroke, and presence of subarachnoid contrast. Cohen's κ statistic was used to assess interrater agreement for each imaging outcome and compared the FDCT to conventional imaging.
Results:
FDCT for ICH demonstrated a strong degree of interrater reliability (κ = 0.896; 95% confidence interval [CI], 0.734-1.057). Negligible reliability was seen for ischemia determination on immediate post-MT FDCT (κ = 0.149; 95% CI, -0.243 to 0.541). ICH evaluation between FDCT and post-MT conventional CT revealed modest interrater reliability (κ = 0.432; 95% CI, -0.100 to 0.965), which did not reach statistical significance. There was no substantive reliability in the evaluation of ICH between FDCT and post-MT MRI (κ = 0.118, 95% CI, -0.345 to 0.580).
Conclusion:
FDCT, such as Dyna CT, immediately post-MT is a promising tool that can expedite the detection of ICH with a high degree of reliability, although the detection of ischemic parenchymal changes is limited.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...