Score Function for the Optimization of the Performance of Forward Fill/Flush Differential Flow Modulation for
Aleksandra Lelevic1,2, Christophe Geantet2, Chantal Lorentz2
1IFP Energies nouvelles, Rond-point de l'échangeur de Solaize BP 3, 69360 Solaize, France.
Abstract:
Modulation is the key element of the comprehensive 2D gas chromatography separation. Forward fill/flush flow modulation is cost effective, robust and suitable for analysis of a wide range of samples. Even though this modulation system is well known, studies regarding its optimization are sparse. In this work, based on hundreds of experiments involving multiple column sets and modulation conditions, an approach was proposed that permits to facilitate the choice of the forward fill/flush flow modulation parameters. A score function was developed that allows to predict the forward fill/flush flow modulation process efficiency as judged by the modulated peak shape. The score function was based on the physical rules for optimized and quantitative forward fill/flush flow modulation proposed in our previous work, which state that the sum of the fill and flush modulation distances should be close to the modulation channel length and that the ratio of the flush and fill distances should be sufficiently high for efficient channel flushing. The score function was embedded in a freely available tool in the form of a forward fill/flush flow modulation calculator, which allows the user either to quickly check the relevancy of the modulation operating conditions or to obtain a suggestion for optimal modulation parameters.
More Related Videos
Related Concept Videos
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Diffusion on Chromatography Columns
Longitudinal diffusion occurs when the solute molecules in the mobile phase diffuse from the more concentrated center of the chromatographic band to the more dilute regions on either side, both towards and against the flow direction. This...