Laser Target System in Combination with an Aiming Device for Percutaneous CT-Guided Interventions - An Accuracy Study
Yannick Scharll1, Daniel Böhler1, Gregor Laimer1
1Interventional Oncology-Microinvasive Therapy (SIP), Department of Radiology, Medical University Innsbruck, Anichstr. 35, 6020 Innsbruck, Austria (Y.S., D.B., G.L., P.S., R.B.).
An aiming device significantly improves laser-guided computed tomography (CT) puncture accuracy in vitro. This tool enhances needle placement precision for CT interventions, offering better targeting in medical procedures.
Area of Science:
- Medical Imaging
- Interventional Radiology
- Surgical Technology
Background:
- Computed tomography (CT)-guided interventions are crucial for accurate diagnoses and treatments.
- Laser guidance systems offer potential for improved targeting in CT procedures.
- Evaluating adjunct devices is essential for optimizing interventional accuracy.
Purpose of the Study:
- To assess the impact of an aiming device on the targeting accuracy of laser-guided punctures during CT interventions.
- To compare needle placement accuracy and procedural time with and without an aiming device.
- To evaluate the effectiveness of combined laser guidance and an aiming device in vitro.
Main Methods:
- 600 CT-guided punctures were performed on conically shaped targets in a plexiglass phantom.
- Procedures utilized a laser target system, with half including an additional aiming device.
- Needle placement accuracy (Euclidean and normal distances) and procedural time were measured across varying CT slice thicknesses (1.25, 2.5, 5 mm).
Main Results:
- The aiming device significantly improved targeting accuracy (p < 0.001) at 1.25 mm and 2.5 mm slice thicknesses.
- Mean normal distances with the aiming device were 1.76 mm (1.25 mm slices), 2.09 mm (2.5 mm slices), and 1.93 mm (5 mm slices).
- Procedural time increased slightly with the aiming device (29.8 min vs. 24.8 min without).
Conclusions:
- The use of an aiming device in conjunction with laser guidance substantially enhances targeting accuracy in CT-guided punctures.
- This combination offers a significant improvement over freehand passes for in vitro CT interventions.
- The aiming device is a valuable adjunct for improving precision in laser-guided CT procedures.