Quantification of Oculomotor Responses and Accommodation Through Instrumentation and Analysis Toolboxes
Published on: March 3, 2023
Reliability and Agreement of an Integrated Platform for Intelligent Visual Function Measurement
Wei-Jun Cai1, Sisi Lin2, Ruru Chen1
1School of Ophthalmology and Optometry and Eye Hospital, Wenzhou Medical University, Wenzhou, Zhejiang, China.
The new Inspection Platform of Visual Function (IPVF) demonstrates high reliability and good agreement with conventional phoropters for visual function assessment, showing excellent repeatability for phoria, negative/positive relative accommodation (NRA/PRA), and accommodative amplitude (AMP).
Area of Science:
- Ophthalmology and Optometry
- Vision Science
- Clinical Instrumentation
Background:
- Phoropters are standard clinical tools for refraction and visual function testing.
- Assessing the reliability of new visual function assessment technologies is crucial for clinical adoption.
Purpose of the Study:
- To evaluate the reliability and agreement of the new Inspection Platform of Visual Function (IPVF) compared to a conventional phoropter (TOPCON VT-10).
- To determine the repeatability and concordance of key visual function measurements between the IPVF and phoropter.
Main Methods:
- A prospective study involving 80 eyes of 80 healthy subjects.
- Measurements included horizontal phoria (Phoria_D, Phoria_N), negative/positive relative accommodation (NRA/PRA), and accommodative amplitude (AMP).
- Intraclass correlation coefficients (ICCs) assessed repeatability; Bland-Altman plots evaluated agreement between the IPVF and phoropter.
Main Results:
- Both IPVF and phoropter showed high repeatability for phoria and accommodative amplitude (ICCs 0.87-0.98).
- IPVF demonstrated excellent repeatability for NRA/PRA (ICC 0.96), while phoropter showed acceptable repeatability (ICC 0.732).
- Narrow 95% limits of agreement indicated satisfactory concordance between the two instruments for all measured parameters.
Conclusions:
- The IPVF instrument exhibits high repeatability and satisfactory agreement with a conventional phoropter for visual function assessment.
- The IPVF showed slightly superior repeatability for PRA compared to the phoropter.
- The findings support the IPVF as a reliable tool for clinical visual function testing.