Dual-Tracer Parathyroid Imaging Using Joint SPECT Reconstruction
Jaruwan Onwanna1,2, Maythinee Chantadisai3, Tawatchai Chaiwatanarat2,4
1Biomedical Engineering Program, Faculty of Engineering, Chulalongkorn University, Bangkok, Thailand.
Purpose:
We assessed the lesion detection performance of the dual-tracer parathyroid SPECT imaging using the joint reconstruction method.
Materials And Methods:
Thirty-six noise realizations were created from SPECT projections collected from an in-house neck phantom to emulate 99mTc-pertechnetate/99mTc-sestamibi parathyroid SPECT datasets. Difference images representing parathyroid lesions were reconstructed using the subtraction and the joint methods whose corresponding optimal iteration was defined as the iteration which maximized the channelized Hotelling observer signal-to-noise ratio (CHO-SNR). The joint method whose initial estimate was derived from the subtraction method at optimal iteration (the joint-AltInt method) was also assessed. In a study of 36 patients, a human-observer lesion-detection study was performed using difference images from the three methods at optimal iteration and the subtraction method with four iterations. The area under the receiver operating characteristic curve (AUC) was calculated for each method.
Results:
In the phantom study, both the joint-AltInt method and the joint method improved SNR compared to the subtraction method at their optimal iteration by 444% and 81%, respectively. In the patient study, the joint-AltInt method yielded the highest AUC of 0.73 as compared with 0.72, 0.71, and 0.64 from the joint method, the subtraction method at optimal iteration, and the subtraction method at four iterations. At a specificity of at least 0.70, the joint-AltInt method yielded significantly higher sensitivity than the other methods (0.60 vs 0.46, 042, and 0.42; p < 0.05).
Conclusions:
The joint reconstruction method yielded higher lesion detectability than the conventional method and holds promise for dual-tracer parathyroid SPECT imaging.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
07:13Author Spotlight: An Efficient and Robust Software for Automated Fusion of Multiple Preclinical Imaging Modalities
Published on: October 27, 2023
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT