Dosimetry for Cell Irradiation using Orthovoltage 40-300 kV X-Ray Facilities
Published on: February 20, 2021
Radiosurgery dosimetry using CaSO4:Eu OSLD film-a feasibility study
Patricia Br Gasparian1,2, Anna Lmc Malthez3, Michele Torquato4
1Comissão Nacional de Energia Nuclear-CNEN, Rio de Janeiro, Brazil.
This study introduces a novel, reusable optically stimulated luminescence dosemeter (OSLD) film for 2D dose mapping. The CaSO4:Eu-based film offers a low-cost, simple method for radiosurgery dose distribution analysis.
Area of Science:
- Medical Physics
- Radiation Dosimetry
- Materials Science
Background:
- Optically stimulated luminescence (OSL) systems offer fast and simple 2D dose evaluation with submillimeter resolution.
- Existing OSL methods may lack cost-effectiveness or reusability for widespread application.
Purpose of the Study:
- To develop and characterize a novel, reusable optically stimulated luminescence dosemeter (OSLD) film for 2D dose mapping.
- To assess the dosimetric properties and feasibility of CaSO4:Eu embedded in a silicone elastomer matrix for radiation therapy applications.
Main Methods:
- Fabrication of a CaSO4:Eu-embedded silicone elastomer film using a low-cost methodology.
- Evaluation of dosimetric properties including repeatability, homogeneity, dose response, and signal fading using a TL/OSL Risø reader.
- Application of the OSLD film for 2D dose distribution mapping in a 6 MV photon beam for radiosurgery.
Main Results:
- The reusable OSLD film demonstrated repeatability within 3% and linearity from 5 to 25 Gy.
- Observed a sensitivity change of nearly 12% within a 5x5 cm2 film, indicating room for homogeneity improvement.
- Reported significant OSL signal fading (approx. 50% in the first week), followed by stabilization.
- Successfully mapped dose distribution in radiosurgery, proving the film's practical utility.
Conclusions:
- The developed CaSO4:Eu OSLD film presents a feasible, reusable, and low-cost solution for 2D dosimetry.
- The film's performance in mapping dose distribution in radiosurgery highlights its potential in clinical radiation oncology.
- Further optimization for homogeneity and fading characteristics could enhance the OSLD film's applicability.