Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
CT Number Accuracy and Association With Object Size: A Phantom Study Comparing Energy-Integrating Detector CT and
Aria M Salyapongse1, Sean D Rose2, Perry J Pickhardt1,3
1Department of Radiology, University of Wisconsin Madison, 1005 Wisconsin Institute for Medical Research, 1111 Highland Ave, Madison, WI 53705.
Photon-counting detector (PCD) CT offers more reliable CT numbers than energy-integrating detector (EID) CT, showing less variation with patient size. This advancement in CT technology improves accuracy for diagnostic imaging.
Area of Science:
- Medical Imaging
- Radiological Physics
- Detector Technology
Background:
- Energy-integrating detector (EID) CT exhibits variable CT numbers due to beam hardening influenced by patient size.
- Photon-counting detector (PCD) CT provides more accurate effective beam energy determination, enhancing CT number reliability.
Purpose of the Study:
- To compare deep silicon PCD CT with EID CT (single- and dual-energy modes).
- To evaluate the impact of object size on CT numbers.
- To assess the overall accuracy of CT numbers across different detector types.
Main Methods:
- Scanned a phantom with varying polyethylene ring sizes and material inserts using EID CT (SE and DE modes) and PCD CT.
- Measured CT numbers in regions of interest (ROIs) and computed slopes of CT number versus object size.
- Calculated ideal CT numbers using the NIST XCOM database and determined the root mean square error (RMSE) for accuracy assessment.
Main Results:
- PCD CT demonstrated significantly lower CT number variation with object size (slopes closer to zero) compared to both SE and DE EID CT for various materials.
- PCD CT exhibited the smallest RMSE across all tested materials, indicating CT numbers closest to ideal values.
- Specific RMSE values showed PCD CT (17 HU for air, 3 HU for water, 4 HU for polystyrene, 13 HU for iodine, 20 HU for bone) outperformed EID CT (SE: 32, 7, 9, 31, 69 HU; DE: 44, 8, 10, 37, 81 HU).
Conclusions:
- Deep silicon PCD CT offers superior performance over SE and DE EID CT regarding CT number stability with varying object sizes.
- PCD CT provides CT numbers with higher accuracy, closer to ideal values, compared to EID CT methods.
- The enhanced reliability of CT numbers from PCD CT is crucial for diagnostic pathways reliant on quantitative imaging data.