Related Experiment Video
Updated: Jul 25, 2025

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Characterization of a new radiochromic film (LD-V1) using mammographic beam qualities
Erika Nakajima1, Hitoshi Sato1
1Department of Radiological Sciences, Ibaraki Prefectural University of Health Sciences, Ibaraki Prefecture, Japan.
Purpose:
Radiochromic film (RCF) is a detector that can obtain a two-dimensional dose distribution with high resolution; it is widely used in medical and industrial fields. Several types of RCFs exist based on their application. The type of RCF mainly used for mammography dose assessment has been discontinued; however, a new type of RCF (LD-V1) has been distributed as a successor. Since the medical use of LD-V1 has rarely been studied, we investigated the response characteristics of LD-V1 in mammography.
Methods:
Measurements were performed using Mo/Mo and Rh/Ag on a Senographe Pristina mammography device (GE, Fairfield, CT, USA). The reference air kerma was measured using a parallel-plate ionization chamber (PPIC) (C-MA, Applied Engineering Inc, Tokyo, Japan). Pieces of LD-V1 film model were irradiated at the same position where the reference air kerma in air was measured by the PPIC. Irradiation was performed using the time scale method based on the load on the equipment. Two methods of irradiation were considered: placing the detector in air and on the phantom. The LD-V1 was scanned five times at 72 dpi in RGB (48 bit) mode using a flatbed scanner (ES-G11000, Seiko Epson Corp, Nagano, Japan) 24 h following irradiation. The response ratio of the reference air kerma and the air kerma obtained from the LD-V1 were compared and examined for each beam quality and air kerma range.
Results And Discussion:
When the beam quality was altered, the response ratio varied from 0.8 to 1.2 with respect to the measurement value of the PPIC; however, some outliers were observed. Response ratios were highly variable in the low-dose range; however, as the air kerma increased, the ratios approached 1. Thus, LD-V1 does not need calibration for each beam quality used in mammography. LD-V1 enables air kerma evaluation by creating air kerma response curves under certain X-ray conditions used in mammography.
Conclusion:
We suggest that the dose range be limited to 12 mGy or more to keep the response variation with beam qualities below ±20%. If further measurement is required for reducing the response variation, the dose range should be shifted to a higher dose range.
Related Concept Videos
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation I: X-ray and CT

