Related Experiment Video
Updated: Jul 23, 2025

An Inertial Measurement Unit Based Method to Estimate Hip and Knee Joint Kinematics in Team Sport Athletes on the Field
Published on: May 26, 2020
Monitoring Knee Contact Force with Force-Sensing Insoles
Alex Spencer1,2,3, Michael Samaan2,3,4, Brian Noehren1,4
1Department of Physical Therapy, College of Health Sciences, University of Kentucky, Lexington, KY 40508, USA.
Abstract:
Numerous applications exist for monitoring knee contact force (KCF) throughout activities of daily living. However, the ability to estimate these forces is restricted to a laboratory setting. The purposes of this study are to develop KCF metric estimation models and explore the feasibility of monitoring KCF metrics via surrogate measures derived from force-sensing insole data. Nine healthy subjects (3F, age 27 ± 5 years, mass 74.8 ± 11.8 kg, height 1.7 ± 0.08 m) walked at multiple speeds (0.8-1.6 m/s) on an instrumented treadmill. Thirteen insole force features were calculated as potential predictors of peak KCF and KCF impulse per step, estimated with musculoskeletal modeling. The error was calculated with median symmetric accuracy. Pearson product-moment correlation coefficients defined the relationship between variables. Models develop per-limb demonstrated lower prediction error than those developed per-subject (KCF impulse: 2.2% vs 3.4%; peak KCF: 3.50% vs. 6.5%, respectively). Many insole features are moderately to strongly associated with peak KCF, but not KCF impulse across the group. We present methods to directly estimate and monitor changes in KCF using instrumented insoles. Our results carry promising implications for internal tissue loads monitoring outside of a laboratory with wearable sensors.