Development and multi-institutional evaluation of a new phantom for verifying beam-positioning errors at
Takahisa Kido1, Tomohiro Ono2, Mitsuhiro Nakamura1
1Department of Advanced Medical Physics, Graduate School of Medicine, Kyoto University, Kyoto, Japan.
A new phantom accurately measures beam positioning errors in single-isocenter stereotactic radiotherapy for brain metastases. This tool ensures precise off-isocenter delivery, crucial for treating multiple brain lesions effectively.
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiotherapy Technology
Background:
- Single-isocenter stereotactic radiotherapy is essential for treating multiple brain metastases.
- Accurate delivery at off-isocenter positions is critical for treatment efficacy and patient safety.
Purpose of the Study:
- To verify beam-positioning errors at off-isocenter locations using a novel phantom.
- To assess the accuracy and reproducibility of treatment delivery in single-isocenter stereotactic radiotherapy.
Main Methods:
- A phantom with five stainless-steel balls was designed to evaluate off-isocenter beam positioning.
- Phantom rotations (0° and 90°) and multi-institutional testing on eight linacs were performed.
- Reproducibility was assessed via standard deviation (SD) at a single institution.
Main Results:
- The phantom setup and image acquisition took approximately 20 minutes.
- Reproducibility tests showed a standard deviation of 0.41 mm for beam-positioning errors.
- Multi-institutional evaluation revealed errors within 1.00 mm of tolerance at the isocenter, with the largest error (1.36 mm) observed 7.5 cm off-isocenter.
Conclusions:
- The developed phantom is a valuable tool for establishing beam-positioning errors in single-isocenter stereotactic radiotherapy.
- It facilitates accurate assessment of off-isocenter treatment delivery, enhancing safety for patients with brain metastases.