A novel esophageal retractor with eccentric balloon during atrial fibrillation ablation
Wen-Li Dai1, Ke-Xin Yao1, Meng-Meng Li1
1Department of Cardiology, National Clinical Research Center for Cardiovascular Diseases, Beijing Anzhen Hospital, Capital Medical University, Chaoyang-qu, Beijing, China.
Background:
Due to the anatomically adjacent relationship between the left atrium (LA) and esophagus, energy delivery on the posterior wall of LA is limited. The aim of this study was to evaluate the feasibility of a novel esophageal retractor (SAFER) with an inflatable C-curve balloon during atrial fibrillation (AF) ablation.
Method:
Nine patients underwent AF ablation assisted with the SAFER. After inflation, the esophagus was deviated laterally away from the intended ablation site of the posterior wall under local anesthesia. The extent of mechanical esophageal deviation (MED) was evaluated under fluoroscopy, defined as the shortest distance from the trailing esophageal edge to the closest point of the ablation line. Gastroscopy was performed before and after ablation. The target ablation index used in all LA sites including the posterior wall was 400-450 after effective MED. All adverse events during the periprocedural period were recorded.
Results:
The mean deviation distance achieved 16.2 ± 9.6 mm away from the closest ablation point of the pulmonary vein lesion set. With respect to the individual left and right pulmonary vein lesion sets, the deviation distance was 19.7 ± 11.5 and 12.7 ± 6.8 mm, respectively. The extent of deviation was 0 to 5 mm, 5.1 to 10 mm, or >10 mm in 0(0%), 7(38.9%), and 11(61.1%), respectively. Procedural success was achieved in all patients without acute reconnection. There was only one esophageal complication which manifested as esophageal erosion and this patient experienced throat pain possibly related to the SAFER retractor with no clinical sequelae.
Conclusion:
Esophageal deviation with the novel eccentric balloon is a novel feasible choice during AF ablation, enabling adequate energy delivery to the posterior wall of LA. Additional prospective randomized controlled studies are required for further validation.